Managing Patients
2000966-386D
Dash™ 3000/4000/5000
5-9
Option
Software Option
Function
VIEW ON ALARM
All
1
Select one of the following options:
OFF
— Do not allow this monitor to view other
monitors.
ON
— Display alarm messages from alarming monitors
in the care unit.
If running AutoView XM, the alarm levels that trigger an
automatic display on a host monitor are defined in the
AVOA ALARM LEVEL
setting.
AUTO
— When running AutoView XM, select this
option to display alarm messages and enable an
automatic split-screen view on alarm on a host monitor.
VOA BROADCAST
All
Select one of the following options:
OFF
— Do not send any alarms from this monitor to
other monitors on the network.
ON
— Send all
ADVISORY
,
WARNING
and
CRISIS
alarms to other monitors on the network. Only monitors
that have this monitor selected to automatically view on
alarm receive the alarms.
WARNING
ALARM NOTIFICATION LIMITATION—If a viewing bed is
configured to auto view a bed that has
SEND AUTO VIEWS
turned
OFF
, the viewing bed will only receive the alarm line
message, NOT the split-screen view or the accompanying
audible alarm. To avoid this limitation, verify that all viewing
beds in the watch list have
SEND AUTO VIEWS
turned
ON
.
VOA ALERT TONE
All
Select one of the following options:
OFF
— No tone sounds at host monitor.
ON
— Tone sounds once at host monitor.
REPEAT
— Tone sounds at the host monitor until the
condition is resolved.
AVOA ALARM LEVEL
AutoView XM
Select the lowest alarm level to automatically view on the
host monitor. All alarm levels of higher priority will also
display.
NOTE
If the monitor is not running AutoView XM, all patient
status warnings, system warnings and
CRISIS
alarms
in the care unit display in the alarm message window.
REMOTE SILENCE
All
Select the highest alarm level that can be silenced at the
host monitor. All alarm levels of lesser priority can also be
silenced. Alarms cannot be silenced in
OPERATING
ROOM
.
AVOA SILENCE
AutoView XM
Select when the split view of the alarming monitor is
removed after an alarm is silenced at the host monitor.
SILENCED
— Split view of silenced monitor displays
until
RETURN
or
MAIN MENU
is selected, even if
another monitor is alarming.
NEW ALARM
— Split view of silenced monitor
replaced when another monitor alarms.
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......