5-6
Dash™ 3000/4000/5000
2000966-386D
Managing Patients
Discharge
Discharging a patient from a monitor clears all patient information while retaining the
unit name, bed number and factory or custom defaults.
WARNING
DISCHARGE TO CLEAR PATIENT DATA—When admitting a
new patient to the monitor, you must clear all previous patient data
from the monitor. To accomplish this, disconnect all patient cables
and discharge the patient from the monitor.
To manually discharge a patient from the monitor, follow this procedure:
1.
Disconnect all patient cables.
2.
Select
MORE MENUS
>
ADMIT MENU
>
DISCHARGE PATIENT
or
NEW
CASE
.
3.
If in
COMBO
or
ROVER
COMBO
monitoring modes, select on of the following
to terminate telemetry monitoring:
MONITOR
— Discharge from monitor only. If ECG was acquired from
telemetry, telemetry will be monitored by the central station.
TELEMETRY
— Discharge from telemetry only. ECG monitoring
automatically switches to the monitor.
BOTH
— Discharge from monitor and telemetry.
4.
If the monitor is setup to view other monitors, the following events occur when
the host monitor is discharged:
When running the VOA or AVOA software option, the viewed bed is
removed from the display and the monitor reverts to the factory or custom
defaults.
When running the AutoView XM software option, the viewed bed is
removed from the display, the monitor reverts to the factory or custom
defaults. All monitors from other care units are removed from the list of
watched monitors. All monitors in the same care unit are retained and
monitored according to the
VIEW ON ALARM
control settings.
SET LOCATION ID
ROVER
and
ROVER
COMBO
All
Select the MUSE Cardiology Information System for network
identification.
ECG SOURCE
COMBO
and
ROVER
COMBO
All
Select the monitor or telemetry transmitter to receive ECG data. If
you select telemetry you must also select the transmitter number
before you can admit a patient to the monitor.
1
All includes all four monitoring modes; STANDARD, ROVER, COMBO and ROVER COMBO.
2
All includes all three patient-monitor types; ADULT-ICU, NEONATAL-ICU and OPERATING ROOM.
Option
Monitoring
Mode
Patient-Monitor
Type
Function
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......