![GE Dash 3000 Operator'S Manual Download Page 184](http://html.mh-extra.com/html/ge/dash-3000/dash-3000_operators-manual_128578184.webp)
17-6
Dash™ 3000/4000/5000
2000966-386D
Monitoring Impedance Cardiography (ICG)
Troubleshooting
For clinical-level troubleshooting, refer to the GE “Critical Care Monitoring Clinical
Reference and Troubleshooting Guide.”
SPEED
Select a sweep speed for the displayed waveform.
SIGNAL QUALITY
Select one of the following options for signal quality. This
overrides the automatic signal selection made by the
module for data acquisition and processing.
AUTO SEARCH
— Search all leads for optimum
signal. Displays the message “Signal Search” in the
parameter window.
SIGNAL 1
to
SIGNAL 4
— Select to switch the vector
designation to a specific lead.
BEAT AVERAGE
Select the number of heart beats that are averaged. When
monitoring begins, beat averaging begins at 10, 20, 30 and
60 beats. The averaging remains at 60 beats unless a new
value is selected. The smaller the number selected, the
more likely the data will be affected by artifact. The higher
the number selected, the more the beat average will be
smoothed out with minimum fluctuation.
CHANGE NORMALS
Select the normal range for the parameters.
CHECK LEADS
Check for lead failures.
Option
Function
Message
Cause
Solution
ENTER REQUIRED ICG PATIENT INFO
Required patient information missing or
incorrect.
Enter the required patient information.
ENTER PATIENT WEIGHT
ENTER PATIENT HEIGHT
ENTER PATIENT AGE
ENTER PATIENT SEX
PATIENT WEIGHT OUT OF RANGE
PATIENT HEIGHT OUT OF RANGE
PATIENT AGE OUT OF RANGE
SERVICE MODULE
Hardware failure.
Contact service.
RIGHT THORACIC LEAD FAIL
Lead failure.
Check patient and sensor.
LEFT THORACIC LEAD FAIL
RIGHT NECK LEAD FAIL
LEFT NECK LEAD FAIL
CABLE DISCONNECTED
Hardware or lead failure.
SIGNAL SEARCH
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......