![GE Dash 3000 Operator'S Manual Download Page 167](http://html.mh-extra.com/html/ge/dash-3000/dash-3000_operators-manual_128578167.webp)
Monitoring End-Tidal CO2
2000966-386D
Dash™ 3000/4000/5000
15-5
Calibrating
Sensors
To calibrate a sensor, follow this procedure.
1.
Remove the adapter from the sensor.
Option
Function
UNITS
Select the units of measure of inspired and expired CO
2
.
NOTE
Respiration is measured in breaths per minute; no
breath is measured in seconds.
CO2 SCALE
Select a scale for the displayed CO
2
capnogram. The larger
the scale, the smaller the displayed waveform. A waveform
scale that exceeds the defined display area is flattened on
top.
CO2 LIMITS
Select high and low alarm limits for inspired/expired CO
2
,
respiration rate and apnea. An alarm sounds when a value
occurs outside of these limits.
N20 COMPENSATION
NOTE
Only available when in the
OPERATING ROOM
patient-monitor type.
Select the percentage of N
2
O concentration to compensate
for N
2
O-related elevation of the CO
2
value. Obtain the
concentration from an analyzer or anesthesia machine.
O2 COMPENSATION
Select the percentage of O
2
concentration to compensate
for O
2
-related decrease of the CO
2
value. Obtain the
concentration from an oxygen analyzer.
CO2 AVERAGING
Select a time interval for averaging CO
2
.
CAL SENSOR TO ZERO
CELL
NOTE
Only available with a Mainstream setup.
Zero the sensor each time a sensor is connected to the
monitor.
CALIBRATE SAMPLE
LINE
NOTE
Only available with a Sidestream setup.
Calibrate the module each time a module is connected to
the monitor.
CALIBRATE ADAPTER
NOTE
Only available with a Mainstream setup.
Calibrate the airway adapter each time a adapter is
connected to the monitor.
SPEED
Select the sweep speed for the displayed waveform.
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......