12-4
Dash™ 3000/4000/5000
2000966-386D
Measuring Cardiac Output
Performing the CO Procedure
To perform a CO procedure, follow this procedure. Before you begin this procedure,
verify you are ready by reviewing
1.
Select the
CO
parameter window label.
2.
If the monitor is configured for automatic measurements, the “
INJECT WHEN
READY
” message and
CARDIAC OUTPUT
information window displays.
3.
Watch for the
“Inject Now
” message and select
CARDIAC OUTPUT NOW
to
calculate the cardiac output.
The information window displays the current cardiac trial information, a review
CO curve, the derived CO value, the average CO value between stored trials, the
patient’s heart rate and other vital signs at the time of the trial. It also includes
cardiac disturbances (
RHY
), the patient’s height and weight and calculated
parameters.
PRINT CO CURVE
Print the displayed real-time waveform. A “
PRINTING
MANUAL
” message displays when the waveform is printing.
AUTO MODE
Turn automatic measurements on or off. The monitor
searches for a stable baseline by averaging approximately
8.5 seconds of the patient’s blood temperature and displays
the message “
INJECT WHEN READY
” when a stable
baseline is established.
USE
Before the injection procedure begins, select one of the
blood pressure values to be factored in the cardiac
calculation:
CATHETER
Before the injection procedure begins, select the correct
catheter type. The monitor guides you through the
INJECT
TEMP
,
SIZE
and
INJECT VOL
options to assure correct
setup.
If you select
OTHER
, you must select a value for
COMPUTATION CONSTANT
. Refer to the literature
provided by the catheter manufacturer for the correct value.
INJECT TEMP
Select the injectate probe type.
SIZE
Select the catheter French size.
INJECT VOL
Select the injectate volume.
COMPUTATION
CONSTANT
Select a computation constant for a catheter manufacturer.
If you selected a specific manufacturer for the
CATHETER
,
the correct value for the manufacturer is automatically
selected for you.
If you selected
OTHER
for the
CATHETER
, you must select
a value or you will not be able to complete a trial. Refer to
the literature provided by the catheter manufacturer for the
correct value.
CARDIAC OUTPUT HELP
Display troubleshooting information.
Option
Function
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......