Monitoring SpO2
2000966-386D
Dash™ 3000/4000/5000
11-7
WARNING
INACCURATE MEASUREMENTS—Inaccurate measurements
may be caused by various patient or mechanical conditions. These
conditions include but are not limited to: venous pulsations;
hypotension; severe vasoconstriction; severe anemia; hypothermia;
shock; cardiac arrest; sensor tension; sensor placement on the same
extremity as a blood pressure cuff, arterial catheter or intravascular
line; or arterial occlusion proximal to the sensor.
WARNING
INTERFERING SUBSTANCES—Carboxyhemoglobin may
erroneously increase SpO
2
readings. The level of increase is
approximately equal to the amount of carboxyhemoglobin present.
Dyes, or any substance containing dyes, that change usual arterial
pigmentation may cause erroneous readings.
WARNING
MRI INTERFERENCE—Do not monitor SpO
2
or use SpO
2
probes
during magnetic resonance imaging (MRI) scanning. Induced
current could potentially cause burns. SpO
2
monitoring may affect
the MRI image, and the MRI unit may affect the accuracy of the
oximetry measurements.
NOTE
Read all literature accompanying sensors for specific safety information.
Neonates and Infants
The following precautions apply when monitoring neonate and infant patients.
WARNING
—The display of inaccurate pulse oximetry values has been linked
to the presence of poor signal strength or artifact due to patient
motion during signal analysis. This condition is most likely to be
encountered when the monitor is used on neonates or infants. These
same conditions in adults do not impact the values to the same
extent.
Use the following criteria when measuring SpO
2
on neonates and infants.
The peripheral pulse rate (PPR) as determined by the SpO
2
function must be
within ten percent of the ECG heart rate.
The SpO
2
signal strength indicator must have two or three asterisks displayed.
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......