Monitoring SpO2
2000966-386D
Dash™ 3000/4000/5000
11-3
Primary Parameter Monitoring Considerations
When ECG is turned off, the SpO
2
parameter becomes the primary patient monitoring
parameter. The following conditions apply when ECG is turned off:
The SpO
2
and SpO
2
Rate parameter patient status alarm levels default to
WARNING
.
The SpO
2
Pulse Search and Probe Off system status alarm levels default to
WARNING
.
SpO
2
becomes the
WAVEFORM 1
selection for waveform display and graphing.
Configuration Options
To determine the option running on the monitor, look on the right side of the monitor
next to the patient connectors. There will be a Masimo SET, Nellcor OxiMAX or an
Ohmeda label. This label indicates the technology option installed on the monitor.
NOTE
The patient cable should easily plug into the patient connector. Do not use
excessive force to connect the cable. If the patient cable does not easily fit into
the patient connector, it is likely that you do not have the appropriate cable for the
enabled configuration option.
Masimo
The following measurement guidelines apply to Masimo SET:
The time period for acquiring a measurement average is adjustable.
The default setting for all patient-monitor types is eight seconds.
The pulse oximetry is calibrated to display functional saturation.
The signal strength indicator denotes signal quality. Zero asterisks indicate low
signal quality; three asterisks indicate high signal quality.
Only Masimo LNOP or LNCS sensors are supported.
Masimo LNOP or LNCS sensors non-invasively measure pulse rate and the
amount of oxygenated hemoglobin.
Possession or purchase of this device does not convey any express or implied
license to use the device with unauthorized replacement parts which would,
alone, or in combination with this device, fall within the scope of one or more of
the patents relating to this device.
NOTE
With motion, the plethysmographic waveform is often distorted and may be
obscured by the artifact. With Masimo SET technology, the SpO
2
waveform is
not an indication of signal quality or validity. Even with an SpO
2
waveform
obscured by artifact, Masimo SET technology is able to read through the noise
and locate the arterial pulsation.
Use the following guidelines when using Masimo LNOP or LNCS sensors:
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......