10-10
Dash™ 3000/4000/5000
2000966-386D
Monitoring NBP
To set the system status alarm level, select
MORE MENUS
>
MONITOR SETUP
>
MONITOR DEFAULTS
>
NBP NO DETERMINATION
and select
SYSTEM
ADVISORY
or
SYSTEM WARNING
. This monitor default controls both no
determination and pump timeout.
Extended NBP Alarm Silence
To extend the alarm silence, follow this procedure.
1.
Discharge the patient.
2.
Select
MORE MENUS
>
MONITOR SETUP
>
MONITOR DEFAULTS
>
SETUP DEFAULT DISPLAY
.
3.
From the
SETUP DISPLAY
information window, locate the
NBP SILENCE
ALARM
option and
SELECT > 1 MINUTE
.
4.
Press the
Silence Alarm/Admit
key.
The NBP alarm is silenced until you change the
NBP SILENCE ALARM
setting
to
NORMAL
or discharge the patient.
All patient status alarms that occur while alarms are silenced are stored in
ALARM
HISTORY
(as a patient status
MESSAGE
).
Troubleshooting
For clinical-level troubleshooting, refer to the GE “Critical Care Monitoring Clinical
Reference and Troubleshooting Guide.”
NOTE
The following messages display until the next measurement is initiated or
manually cleared.
Problem
Monitor Response
Solution
CUFF INFLATION OVER 5 MINS
Stops automatic NBP measurement after
ONE message.
Make sure there is no air in the cuff and try a
manual measurement.
DEFLATION FAIL, REMOVE CUFF
Remove cuff and contact service.
NBP HARDWARE MALFUNCTION
Contact service.
OVER PRESSURE
Remove cuff and contact service.
NO DETERMINATION
Stops automatic NBP measurement after
TWO messages.
Verify cuff placement and try a manual
measurement.
PUMP TIMEOUT
(Inflation failure or pressure leak.)
Check connections between cuff and
monitor, check for worn connector “O” rings
and try a manual measurement.
TOTAL TIMEOUT
(No measurement in over three minutes for
adult and pediatrics or 90 seconds for
neonatal patients.)
Verify patient is not moving excessively or
experiencing an arrhythmia condition and
try a manual measurement.
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......