Monitoring NBP
2000966-386D
Dash™ 3000/4000/5000
10-3
NOTE
Values change to
X
when no monitoring has taken place for more than 15
minutes (
OPERATING ROOM
), two hours (
ADULT-ICU
) or 12 hours
(
NEONATAL-ICU
).
Options
The monitor supports two blood pressure technologies. To determine the technology
on this monitor, select
MORE MENUS
>
MONITOR SETUP
>
SOFTWARE
CONFIGURATION
. The technology is listed under
NBP
.
DINAMAP Classic
CAUTION
NBP READINGS MAY TIME OUT WHEN USING IABP—An
IABP balloon pump creates non-physiologic arterial waveforms.
These waveforms create an oscillometric signal that may not be
interpreted by DINAMAP Classic, causing NBP to time out.
The patient blood pressure can be monitored from the balloon pump
device.
The first determination initially pumps up to a target cuff pressure of about 160
mmHg for adult/pediatric patients or 110 mmHg for neonates depending on initial
target pressure preset.
DINAMAP SuperSTAT
The first determination initially pumps up to a target cuff pressure of about 135
mmHg for adults or 110 mmHg for neonates, depending on initial target pressure
preset. To allow for rapid setting of cuff pressure, the monitor will momentarily
inflate to a higher pressure, then immediately deflate to the target pressure.
As a determination is taken, the monitor stores the pattern of the patient's oscillation
size as a function of pressure. In any subsequent determination, as few as four
pressure steps may be necessary to complete the process. When employing fewer
pressure steps, the system uses the stored information from the previous blood
pressure determination to decide the best pressure steps to take. The monitor
measures the consistency of pulse size to tell if the oscillations taken at a step are
good and if more steps are needed.
If the current blood pressure reading is similar to the previous reading, the monitor
may use some information from the previous blood pressure in the current
determination. The monitor constantly evaluates data during a measurement and tries
to perform a blood pressure determination in the shortest possible time, providing
greater comfort to the patient.
If it has been 16 minutes or less since the last determination and the current blood
pressure is similar to the previous reading, the monitor will try to make an accelerated
determination of blood pressure.
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......