![GE Dash 3000 Operator'S Manual Download Page 113](http://html.mh-extra.com/html/ge/dash-3000/dash-3000_operators-manual_128578113.webp)
Monitoring ECG
2000966-386D
Dash™ 3000/4000/5000
8-17
NOTE
Upon initialization of ST analysis, regardless of which
IDENTIFY V LEAD
option is selected, the V lead’s ST alarm limits will always be the default limits
set for V1.
When the identified V lead’s position or label is changed, the previous identified
V lead’s ST alarm limits will be retained.
For example, if you change the V lead’s ST alarm limits to +/- 3 and then change
the identified V lead’s position and label, the ST alarm limits will remain at +/- 3
regardless of the default ST alarm limits for the new identified V lead.
This functionality occurs in both hard-wired and
COMBO
mode during 5-lead
ECG monitoring.
Performing 12 Lead Analysis
To analyze 12-lead ECG data, follow this procedure.
IDENTIFY V LEAD
Locate the V lead being used for ECG and ST analysis.
For 5-Leadwire Cables — Verify the V Lead used in ST and
arrhythmia analysis is selected. This V Lead is also used for
ST trends and complex displays.
For 10-Leadwire Cables — Since ST measurement is
performed with a different V lead I than arrhythmia analysis,
any changes made to this selection only apply to the V lead
used for arrhythmia analysis. This V lead is not used for ST
trend and complex displays.
NOTE
Changing the V lead changes the displayed waveform,
analog output, network data, ST trend and template,
graphing and histories.
TREND SCALE
Select a ST trend scale.
ST DISPLAY
Select up to three leads for trend or complex display.
INITIATE 12 LEAD ON ST
HISTORY
Enable automatic initiation of 12 lead analysis when an ST
event is stored in history.
ADJUST ST POINT
Change ST measurement point;
J + 0 MS
to
J + 80 MS
.
Option
Function
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......