Revision C
250cx Series Maternal/Fetal Monitor
7-3
2036946-001
Maternal Heart/Pulse Rate Monitoring: MHR/P Source
MHR/P Source
The MHR/P can be determined by the MECG, MSpO
2
, and NIBP sections of the
monitor. However, the data from only one parameter is:
referred to as the MHR/P source
displayed in the MHR/P area;
used to evaluate an MHR/P alarm condition; and
used to generate the MHR/P trace on the strip chart paper.
1
The source is:
selected via the MHR/P Setup screen
may be manually selected or automatically selected by the monitor according to
the following priority order:
MECG (values updated continuously)
MSpO
2
(values updated continuously)
NIBP (NIBP is available only as an
Auto
selection; the
Manual
selection is
disabled. Values updated only when NIBP determinations are taken)
IMPORTANT:
MSpO
2
AS AN MHR/P SOURCE—If MSpO
2
is selected as the
MHR/P source, an MHR/P alarm only occurs if the pulse rate
value derived from the MSpO
2
sensor violates an MHR/P alarm
limit. The MHR/P values derived from the MECG and NIBP
sections of the monitor are ignored. The heart rate tone varies in
pitch to reflect changes in the maternal oxygen saturation reading.
The pitch rises as the saturation values increase, and lowers as the
saturation values decrease. The pulse rate trend is a grey line
annotated by
MSpO
2
P
.
1
If NIBP is selected as the MHR/P source, there is no trending of the data since these are static
measurements.
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 17: ...Revision C 250cx Series Maternal Fetal Monitor 1 1 2036946 001 1 Safety...
Page 18: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety For your notes...
Page 28: ...1 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety Equipment Symbols...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036946 001 2 Introduction...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Introduction For your notes...
Page 53: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036946 001 4 Setup Procedures...
Page 54: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Setup Procedures For your notes...
Page 65: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036946 001 5 Fetal Heart Rate Monitoring...
Page 77: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036946 001 6 Uterine Activity Monitoring...
Page 117: ...Revision C 250cx Series Maternal Fetal Monitor 10 1 2036946 001 10 Alarms...
Page 118: ...10 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Alarms For your notes...
Page 127: ...Revision C 250cx Series Maternal Fetal Monitor 11 1 2036946 001 11 Recorder Modes...
Page 128: ...11 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Recorder Modes For your notes...
Page 153: ...Revision C 250cx Series Maternal Fetal Monitor 13 1 2036946 001 13 Heartbeat Coincidence...
Page 159: ...Revision C 250cx Series Maternal Fetal Monitor 14 1 2036946 001 14 Waveforms...
Page 160: ...14 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Waveforms For your notes...
Page 165: ...Revision C 250cx Series Maternal Fetal Monitor 15 1 2036946 001 15 Maintenance...
Page 166: ...15 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance For your notes...
Page 174: ...16 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Troubleshooting For your notes...
Page 200: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults For your notes...
Page 212: ...B 6 250cx Series Maternal Fetal Monitor Revision C 2036946 001...
Page 214: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts For your notes...
Page 236: ...D 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Frequently Asked Questions...
Page 249: ......