Spectra Alerts: Alert Indications
Alert Indications
Active Alerts
When the system detects an alert condition, visual and audible indications are
provided. (There is no printed indication of the alert.) The alert level, indicated by
asterisks, flashes and displays in inverse video between the FHR1 and FHR2 areas;
in addition, the associated FHR numerics flash. The audio indication is a pattern of
beeps representing the alert level:
Level 1: Beep___Beep___Beep___ . . .
Level 2: BeepBeep___BeepBeep___BeepBeep___ . . .
Level 3: BeepBeepBeep___BeepBeepBeep___BeepBeepBeep___ . . .
“Beep” represents audio tone sets and “___” is the pause between sets.
The following figure provides an example of an active alert—level one
decelerations.
Example of a Level One Decelerations Alert
Silencing Alerts
Press the
Alarm Silence
button to cancel the audio. The alert level stops flashing and
displays in normal video; however, the associated FHR numerics continue to flash.
For an
active, silenced
alert, the visual indications remain present until the condition
is resolved or the alert is reset. (Refer to “Resetting Alerts” on page C-12.)
138
12
87
98%
MECG
MSpO2
NIBP
15:52:58
US
FECG
TOCO
Freeze
Alarms
Setup
VSHX
25 mm/s
II 2x
MECG
3 UC/10
147
*
For an
active
alert, the FHR numerics
flash, indicating which FHR is associated
with the alert. For a silenced alert, the
numerics continue to flash until the
condition is resolved.
For an
active, unsilenced
alert, the alert
level flashes and displays in inverse
video. After the alert is
silenced
, the alert
level stops flashing and displays in
normal video until the alert is reset or the
condition is resolved.
Alert enabled icon.
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 17: ...Revision C 250cx Series Maternal Fetal Monitor 1 1 2036946 001 1 Safety...
Page 18: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety For your notes...
Page 28: ...1 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety Equipment Symbols...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036946 001 2 Introduction...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Introduction For your notes...
Page 53: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036946 001 4 Setup Procedures...
Page 54: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Setup Procedures For your notes...
Page 65: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036946 001 5 Fetal Heart Rate Monitoring...
Page 77: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036946 001 6 Uterine Activity Monitoring...
Page 117: ...Revision C 250cx Series Maternal Fetal Monitor 10 1 2036946 001 10 Alarms...
Page 118: ...10 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Alarms For your notes...
Page 127: ...Revision C 250cx Series Maternal Fetal Monitor 11 1 2036946 001 11 Recorder Modes...
Page 128: ...11 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Recorder Modes For your notes...
Page 153: ...Revision C 250cx Series Maternal Fetal Monitor 13 1 2036946 001 13 Heartbeat Coincidence...
Page 159: ...Revision C 250cx Series Maternal Fetal Monitor 14 1 2036946 001 14 Waveforms...
Page 160: ...14 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Waveforms For your notes...
Page 165: ...Revision C 250cx Series Maternal Fetal Monitor 15 1 2036946 001 15 Maintenance...
Page 166: ...15 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance For your notes...
Page 174: ...16 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Troubleshooting For your notes...
Page 200: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults For your notes...
Page 212: ...B 6 250cx Series Maternal Fetal Monitor Revision C 2036946 001...
Page 214: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts For your notes...
Page 236: ...D 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Frequently Asked Questions...
Page 249: ......