Maternal Pulse Oximetry Monitoring: Module and Probe Compatibility
Modules and Sensors
The Masimo, Ohmeda, and Nellcor parameters are used to measure the amount of
oxygenated hemoglobin and pulse rate non-invasively. The absorption of selected
wavelengths of light is measured with sensors. Although these parameters process
the MSpO
2
measurements differently, the function and appearance of MSpO
2
on
your monitor is the same.
No Implied License
Possession or purchase of this device does not convey any express or implied license
to use the device with unauthorized replacement parts which would, alone, or in
combination with this device, fall within the scope of one or more of the patents
relating to this device.
Sensors
Before use, carefully read the manufacturer’s sensor directions for use.
CAUTIONS
TISSUE DAMAGE—Tissue damage can be caused by incorrect
application or use of a MSpO
2
sensor, for example by wrapping
the sensor too tightly. Inspect the sensor site as directed in the
sensor’s directions for use to ensure skin integrity and correct
positioning and adhesion of the sensor.
Do not use damaged sensors.
Do not use a sensor with exposed optical components.
Do not re-sterilize single-patient use sensors. For reusable sensors, refer
to the manufacturer’s instructions for cleaning and sterilization.
Do not immerse the patient cable in water, solvents, or cleaning solutions
(the patient cable connectors are not waterproof).
Do not re-sterilize the patient cable by irradiation, steam, or ethylene
oxide.
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 17: ...Revision C 250cx Series Maternal Fetal Monitor 1 1 2036946 001 1 Safety...
Page 18: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety For your notes...
Page 28: ...1 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety Equipment Symbols...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036946 001 2 Introduction...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Introduction For your notes...
Page 53: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036946 001 4 Setup Procedures...
Page 54: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Setup Procedures For your notes...
Page 65: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036946 001 5 Fetal Heart Rate Monitoring...
Page 77: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036946 001 6 Uterine Activity Monitoring...
Page 117: ...Revision C 250cx Series Maternal Fetal Monitor 10 1 2036946 001 10 Alarms...
Page 118: ...10 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Alarms For your notes...
Page 127: ...Revision C 250cx Series Maternal Fetal Monitor 11 1 2036946 001 11 Recorder Modes...
Page 128: ...11 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Recorder Modes For your notes...
Page 153: ...Revision C 250cx Series Maternal Fetal Monitor 13 1 2036946 001 13 Heartbeat Coincidence...
Page 159: ...Revision C 250cx Series Maternal Fetal Monitor 14 1 2036946 001 14 Waveforms...
Page 160: ...14 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Waveforms For your notes...
Page 165: ...Revision C 250cx Series Maternal Fetal Monitor 15 1 2036946 001 15 Maintenance...
Page 166: ...15 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance For your notes...
Page 174: ...16 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Troubleshooting For your notes...
Page 200: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults For your notes...
Page 212: ...B 6 250cx Series Maternal Fetal Monitor Revision C 2036946 001...
Page 214: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts For your notes...
Page 236: ...D 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Frequently Asked Questions...
Page 249: ......