Revision C
250cx Series Maternal/Fetal Monitor
8-11
2036946-001
Maternal Non-Invasive Blood Pressure Monitoring: NIBP Monitoring
Taking a Manual Reading Between Auto Determinations
If the
NIBP Start/Stop
button is pressed during the interval time between automatic
readings, a new determination is initiated.
IMPORTANT:
The countdown timer is
not
reset whenever a manual blood
pressure reading is initiated; the next scheduled automatic
determination will take place as planned.
Venous Return in Auto Mode
When in auto mode, the monitor always waits
at least 30 seconds
from the end of
one blood pressure determination to the beginning of the next. This provides a
minimum time that pressure around the patient’s limb is relieved, to allow for
venous return.
At all settings except 1 minute, if a determination ends with less than 30 seconds
remaining until the next one, that next determination will be cancelled.
NOTE:
The 250cx Series Monitor is factory-set with the optional 1-minute interval
time enabled. For information on disabling the 1-minute interval, refer to
the 250/250cx Series Service Manual.
Example 1
. The auto mode is selected with a time interval of 2 minutes. A
determination begins at 12:00:00. Due to excessive patient movement, the
determination ends at 12:01:35. This leaves only 25 seconds until the next
automatic reading scheduled at 12:02:00. The 12:02:00 determination is
cancelled and the
following
reading will resume at 12:04:00.
The optional 1-minute interval is an exception. When 1 minute is selected, if a
determination ends with less than 30 seconds until the next one, the reading will be
delayed to guarantee 30 seconds between determinations. During the delay,
Wait
appears in the Auto mode timer.
Example 2
. The auto mode is selected with a time interval of 1 minute. An
automatic determination begins at 11:59:00 with the next reading therefore
scheduled for 12:00:00. The 11:59:00 determination ends at 11:59:35. This
leaves only 25 seconds until the next scheduled automatic reading. Instead of
being cancelled, the next reading is reset to start in 30 seconds at 12:00:05. The
additional 5 seconds displays as
Wait
.
Adjusting the Interval Time Between Automatic Determinations
You can adjust the interval time in-between determinations by going back into the
maternal
NIBP Setup
screen. Regardless of whether you are increasing or decreasing
the interval time, the countdown timer resets to the new value. The next automatic
reading will occur after the expiration of the new interval.
Example 1
. The interval time is set at 10 minutes and the countdown timer
shows 4 minutes until the next reading — in other words 6 minutes have
elapsed. If you change the interval time to 15 minutes, the countdown timer will
wait another 15 minutes until the next reading. Therefore a total of 21 minutes
will elapse between readings.
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 17: ...Revision C 250cx Series Maternal Fetal Monitor 1 1 2036946 001 1 Safety...
Page 18: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety For your notes...
Page 28: ...1 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety Equipment Symbols...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036946 001 2 Introduction...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Introduction For your notes...
Page 53: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036946 001 4 Setup Procedures...
Page 54: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Setup Procedures For your notes...
Page 65: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036946 001 5 Fetal Heart Rate Monitoring...
Page 77: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036946 001 6 Uterine Activity Monitoring...
Page 117: ...Revision C 250cx Series Maternal Fetal Monitor 10 1 2036946 001 10 Alarms...
Page 118: ...10 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Alarms For your notes...
Page 127: ...Revision C 250cx Series Maternal Fetal Monitor 11 1 2036946 001 11 Recorder Modes...
Page 128: ...11 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Recorder Modes For your notes...
Page 153: ...Revision C 250cx Series Maternal Fetal Monitor 13 1 2036946 001 13 Heartbeat Coincidence...
Page 159: ...Revision C 250cx Series Maternal Fetal Monitor 14 1 2036946 001 14 Waveforms...
Page 160: ...14 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Waveforms For your notes...
Page 165: ...Revision C 250cx Series Maternal Fetal Monitor 15 1 2036946 001 15 Maintenance...
Page 166: ...15 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance For your notes...
Page 174: ...16 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Troubleshooting For your notes...
Page 200: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults For your notes...
Page 212: ...B 6 250cx Series Maternal Fetal Monitor Revision C 2036946 001...
Page 214: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts For your notes...
Page 236: ...D 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Frequently Asked Questions...
Page 249: ......