8-10
250cx Series Maternal/Fetal Monitor
Revision C
2036946-001
Maternal Non-Invasive Blood Pressure Monitoring: NIBP Monitoring
The systolic and diastolic pressures are each indicated with two or three digits and
separated by a slash (/). The mean arterial pressure is indicated with two or three digits
and enclosed in parentheses. All pressure values are displayed in mmHg or kPa.
Systolic, diastolic, MAP, and pulse rate values are printed on the strip chart paper
annotated by an outlined diamond (
)which marks the time of the reading.
Hydrostatic Effect
If patient is standing, sitting, or inclined ensure that cuffed limb is supported to
maintain cuff at level of patient’s heart. If cuff is not at heart level, the difference in
systolic and diastolic values due to hydrostatic effect must be considered. Add 1.80
mmHg (0.24 kPa) to values for every inch (2.54 cm) above heart level. Subtract 1.80
mmHg (0.24 kPa) from values for every inch (2.54 cm) below heart level.
Manual Mode
In manual mode, press the
NIBP Start/Stop
button to begin a single determination. The
cuff will inflate to the target pressure. If this initial inflation pressure is insufficient,
the unit retries with a higher inflation pressure (+40 mmHg; +5.3 kPa). The
instantaneous cuff pressure is displayed in place of the mean arterial pressure area
and is indicated by the title
Cuff
.
If you have the 256 Monitor–which does not have the NIBP parameter installed–and
press the
NIBP Start/Stop
button, the message
NOT INSTALLED
appears under the
NIBP
label on the monitor’s screen.
Automatic Mode
In auto mode an indefinite series of determinations are made at defined time
intervals. Upon activation, a clock icon (
) displays in the
NIBP
area indicating the
time remaining until the next scheduled automatic determination.
NOTE:
The first automatic determination begins after the
expiration
of one
complete interval time period.
Since the first automatic blood pressure reading will not occur until after a complete
interval time, you may wish to take an initial manual reading by pressing the
NIBP
Start/Stop
push button. Automatic determinations inflate to the target pressure if no
previous values are displayed. If previous values are displayed the cuff inflation
target pressure is based on the previous values. If this initial inflation pressure is
insufficient, the unit retries with a higher inflation pressure. The instantaneous cuff
pressure is displayed in place of the mean arterial pressure area and is indicated by
the title
Cuff
.
WARNING
The NIBP parameter should be set to determine blood pressures
only as frequently as is clinically indicated to ensure adequate
patient monitoring.
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 17: ...Revision C 250cx Series Maternal Fetal Monitor 1 1 2036946 001 1 Safety...
Page 18: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety For your notes...
Page 28: ...1 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety Equipment Symbols...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036946 001 2 Introduction...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Introduction For your notes...
Page 53: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036946 001 4 Setup Procedures...
Page 54: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Setup Procedures For your notes...
Page 65: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036946 001 5 Fetal Heart Rate Monitoring...
Page 77: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036946 001 6 Uterine Activity Monitoring...
Page 117: ...Revision C 250cx Series Maternal Fetal Monitor 10 1 2036946 001 10 Alarms...
Page 118: ...10 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Alarms For your notes...
Page 127: ...Revision C 250cx Series Maternal Fetal Monitor 11 1 2036946 001 11 Recorder Modes...
Page 128: ...11 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Recorder Modes For your notes...
Page 153: ...Revision C 250cx Series Maternal Fetal Monitor 13 1 2036946 001 13 Heartbeat Coincidence...
Page 159: ...Revision C 250cx Series Maternal Fetal Monitor 14 1 2036946 001 14 Waveforms...
Page 160: ...14 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Waveforms For your notes...
Page 165: ...Revision C 250cx Series Maternal Fetal Monitor 15 1 2036946 001 15 Maintenance...
Page 166: ...15 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance For your notes...
Page 174: ...16 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Troubleshooting For your notes...
Page 200: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults For your notes...
Page 212: ...B 6 250cx Series Maternal Fetal Monitor Revision C 2036946 001...
Page 214: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts For your notes...
Page 236: ...D 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Frequently Asked Questions...
Page 249: ......