Revision D
250 Series Maternal/Fetal Monitor
5-49
2020551-001
Troubleshooting:
General Troubleshooting
Internal UA Troubleshooting
MECG Troubleshooting
Table 26. Internal UA Troubleshooting
Problem
Probable Cause
Possible Solution
Internal pressure not measuring correctly.
Transducer not properly connected to
monitor.
Air bubble in dome; or catheter blocked.
Dome is cracked.
Strain gauge not at same height as
catheter tip.
Catheter has fallen out of place.
Catheter or strain gauge not zeroed.
Service required.
Ensure transducer is securely attached
to monitor.
Flush dome and catheter.
Replace dome.
Adjust strain gauge height.
Replace catheter.
Calibrate catheter or strain gauge.
Call Biomedical Engineering
Department.
CHECK IUP
message displayed in UA area
of the display.
Blockage in fluid-filled catheter.
Fetus pressing directly on catheter.
Defective strain gauge or catheter.
Service required.
Flush catheter. Re-zero. Replace
catheter if necessary.
Reposition by twisting catheter.
Replace strain gauge or catheter.
Call Biomedical Engineering
Department.
Table 27. MECG Troubleshooting
Problem
Probable cause
Possible solution
MECG erratic or not functioning properly.
Patient cable not properly connected to
monitor.
Electrodes not properly placed.
Clips not attached to electrodes properly.
Electrode gel dried.
Defective MECG cable.
Service required.
Selected lead providing
inadequate signal.
Ensure patient cable is securely
attached to monitor.
Re-apply electrodes.
Check clip attachments.
Check electrodes and re-apply gel if
necessary.
Replace cable.
Call Biomedical Engineering
Department.
Change lead selection on MHR/P Setup
screen.
Dashes (– – –) shown in MHR/P area of
display.
Monitor unable to make a determination
due to insufficient signal.
Ensure patient is not asystolic.
Ensure electrodes are firmly secured to
patient.
Summary of Contents for Corometrics 250 Series
Page 2: ......
Page 6: ...CE CE ii 0086 ...
Page 14: ...viii 250 Series Maternal Fetal Monitor Revision D 2020551 001 ...
Page 16: ...1 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 29: ...Revision D 250 Series Maternal Fetal Monitor 2 1 2020551 001 2 Equipment Overview ...
Page 30: ...2 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 55: ...3 1 250 Series Maternal Fetal Monitor Revision D 2020551 001 3 Installation ...
Page 56: ...3 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 81: ...Revision D 250 Series Maternal Fetal Monitor 4 1 2020551 001 4 Maintenance ...
Page 82: ...4 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 143: ...Revision D 250 Series Maternal Fetal Monitor 5 1 2020551 001 5 Troubleshooting ...
Page 144: ...5 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 196: ...6 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 239: ...Revision D 250 Series Maternal Fetal Monitor B 1 2020551 001 B Alarms Summary ...
Page 240: ...B 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Alarms Summary For your notes ...
Page 243: ...Revision D 250 Series Maternal Fetal Monitor C 1 2020551 001 C Electromagnetic Compatibility ...
Page 251: ......