5-48
250 Series Maternal/Fetal Monitor
Revision D
2020551-001
Troubleshooting:
General Troubleshooting
External Uterine Activity Troubleshooting
Table 25. External UA Troubleshooting
Problem
Probable Cause
Possible Solution
Tocotransducer not recording contractions.
Transducer not properly connected to
monitor.
Transducer not properly placed.
Transducer not secured to patient.
Defective transducer/cable
assembly.
No maternal contractions.
UA Reference range exceeded.
Ensure that transducer is securely
attached to monitor.
Reposition transducer.
Secure or re-apply transducer to
patient.
Replace transducer/cable
assembly.
Wait.
Loosen belts or remove transducer
from patient. Press
UA Reference
button while no pressure is applied to
transducer button. Re-apply
transducer. Do not overtighten belt.
Press
UA Reference
button again
between contractions.
Flashing “+” sign.
Relative pressure > 100.
Press the
UA Reference
button
between contractions.
CHECK TOCO
message is shown in UA
area of the display area when the
UA
Reference
button is pressed.
UA Reference
button pressed before
UA circuits stabilized.
UA Reference range exceeded due to
over-tightening belt.
Transducer defective.
Service required.
You must wait ten seconds following
powering on the monitor and/or
connecting to the
UA
connector.
Loosen belts or remove transducer
from patient. Press
UA Reference
button while no pressure is applied to
transducer button. Re-apply
transducer. Do not overtighten belt.
Press
UA Reference
button again
between contractions.
Replace transducer.
Call Biomedical Engineering
Department.
Summary of Contents for Corometrics 250 Series
Page 2: ......
Page 6: ...CE CE ii 0086 ...
Page 14: ...viii 250 Series Maternal Fetal Monitor Revision D 2020551 001 ...
Page 16: ...1 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 29: ...Revision D 250 Series Maternal Fetal Monitor 2 1 2020551 001 2 Equipment Overview ...
Page 30: ...2 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 55: ...3 1 250 Series Maternal Fetal Monitor Revision D 2020551 001 3 Installation ...
Page 56: ...3 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 81: ...Revision D 250 Series Maternal Fetal Monitor 4 1 2020551 001 4 Maintenance ...
Page 82: ...4 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 143: ...Revision D 250 Series Maternal Fetal Monitor 5 1 2020551 001 5 Troubleshooting ...
Page 144: ...5 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 196: ...6 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 239: ...Revision D 250 Series Maternal Fetal Monitor B 1 2020551 001 B Alarms Summary ...
Page 240: ...B 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Alarms Summary For your notes ...
Page 243: ...Revision D 250 Series Maternal Fetal Monitor C 1 2020551 001 C Electromagnetic Compatibility ...
Page 251: ......