9-8
170 Series Monitor
Revision C
2000947-004
Maintenance:
Preventative Maintenance Inspection
Unit to Primary Leakage
1. Configure a leakage tester to perform a unit AC line leakage test.
2. Turn ON the monitor using the monitor’s
On/Standby
button.
3. Verify the following:
<300
µ
A
(132–253 VAC, or at the line voltage at your fa10%)
4. Record the primary leakage for the following conditions:
Table 9-1. Unit to Primary Leakage Conditions
Neutral
Ground
Power (Polarity)
__________ Closed
__________ Closed
__________ Normal
__________ Closed
__________ Open
__________ Normal
__________ Closed
__________ Open
__________ Reversed
__________ Closed
__________ Closed
__________ Reversed
__________ Open
__________ Closed
__________ Normal
__________ Open
__________ Closed
__________ Reversed