6-74
CIC Pro™
2026419-033E
Configuration
11. Verify the new settings prior to monitoring patients.
Scenario 2: Add or upgrade a CIC Pro center in a care unit running multiple CIC
Pro centers
When a CIC Pro center is added to or upgraded in a care unit already running other
CIC Pro centers, configuration data common to the care unit is changed in the new or
upgraded CIC Pro center, when it starts up, to match the other CIC Pro centers found
in the care unit network. Configuration data specific only to the new or upgraded CIC
Pro center does not change. Therefore, restore is not useful in this situation. Verify the
new settings prior to monitoring patients.
Scenario 3: Restore multiple CIC Pro centers in a care unit to a previous
configuration
In certain cases, it may be necessary to restore the CIC Pro centers in a care unit to a
previous configuration.
1.
Shut down all but one of the CIC Pro centers in the care unit.
shutdown of the CIC Pro center on page 8-31.
This prevents the restore from being overwritten by associated data from other
CIC Pro centers running on the care unit network.
NOTE
The CIC Pro center application must be stopped in order to perform a
restore. Because configuration data is automatically updated to match other
CIC Pro centers found running in the care unit, if other CIC Pro centers are
left running, when the restored CIC Pro center application starts back up,
some configuration data would be overwritten by the automatic update.
2.
Stop the CIC Pro center application on the one CIC Pro center that was left
running.
a.
From the multi-patient viewer, click
Setup CIC
.
b.
Click the
Service Password
tab.
c.
Type
mms_com
as the password and press
Enter
.
d.
At the Windows command line prompt, type
stop
and press
Enter
.
3.
Disconnect the CIC Pro center from the CARESCAPE Network IX and MC
networks.
4.
Connect a service PC directly to the CIC Pro center and
service interface remotely via the CARESCAPE network IX network on page 4-
7
.
5.
Click
Configuration > Backup/Restore > Unit Defaults
.
Summary of Contents for CIC Pro
Page 12: ...x CIC Pro 2026419 033E ...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction ...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview ...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing ...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing ...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces ...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces ...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation ...
Page 94: ...5 20 CIC Pro 2026419 033E Installation ...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration ...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures ...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures ...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting ...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs ...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive ...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount ...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options ...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance ...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software ...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software ...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software ...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software ...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility ...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules ...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility ...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility ...
Page 313: ......