Hardware Installation
IST 005 (C) 1280 x 1024 Landscape Monitor Interconnections
Page 45
n
IST 005
(C) 1280 x 1024 Landscape Monitor Interconnections
Section 1: Introduction
This job card contains the instructions for interconnecting the Workstation to other equipment and
components as part of the installation. Monitor cabling diagrams are provided to show the
interconnection of one or two landscape monitors to the appropriate video board slot positions at the
rear of the computer housing.
Section 2: Prerequisites
The site must be prepared for installation of the workstation. Refer to the
Centricity RA600
Workstation Technical Reference Manual
for detailed information.
u
The appropriate line voltage, line frequency, and power supply line cords.
u
Ethernet connection.
u
Appropriate furniture for the Workstation (i.e. desk, chair, etc.).
Section 3: Cabling
lists the part numbers of interconnecting cables used during installation.
Refer to
see where various connectors are located on the computer housing. Refer
to included manufacturer’s documentation for access to flat panel connectors.
Caution
!
The maximum magnetic field allowable for installing the Centricity
RA600 Workstation is 1 gauss to ensure proper operation.
Summary of Contents for Centricity RA600
Page 16: ...Introduction Page 16 Illustration 1 2 Computer Unit Front View ...
Page 24: ...Introduction Page 24 Modality Type Description IP Address S M GW Port AE Title ...
Page 206: ...Software Procedures SW 006 Single Media Archive SMA Install Configure Test Page 206 ...
Page 274: ...Renewal Parts Page 274 Jorge Aguirre med ge com Maurice Robert med ge com ...