Revision C
CASE
7-5
2060290-201
Setting Up the Ambulatory Blood Pressure Monitor
Operating Steps
1. Turn the monitor off and on again.
2. Wait for the time to be displayed, then connect the monitor to the PC.
3. Click
Connect
.
The acquisition screen indicates the memory status of the ambulatory
blood pressure monitor.
A: Date and time of the monitor.
B: Measurement protocol.
C: Number of BP measurements and the duration of the protocol.
D: Information whether or not data is stored in the monitor.
With TONOPORT V, version 2.0 and later:
E: Patient ID stored in the TONOPORT device.
F: Patient data from the CASE database assigned to the patient ID.
4. Click the
Clear
Data
button to delete data stored in the monitor (with
TONOPORT V devices, version 2.0 and later, this command will also
clear the patient ID).
Note
If the stored data has not been transferred to the system,
download it as described in the next section.
5. Click
Programming
to select the protocol.
182A-BP
A
B
C
D
E
F
Summary of Contents for CASE V6.7
Page 1: ...GE Healthcare CASE V6 7 Software Version 6 73 Operator s Manual 2060290 201 ENG Revision C ...
Page 4: ...CE 2 CASE Revision C 2060290 201 CE Marking Information For your notes ...
Page 15: ...Revision C CASE 1 1 2060290 201 1 Introduction ...
Page 16: ...1 2 CASE Revision C 2060290 201 For your notes ...
Page 30: ...1 16 CASE Revision C 2060290 201 Service Information ...
Page 31: ...Revision C CASE 2 1 2060290 201 2 Equipment Overview ...
Page 32: ...2 2 CASE Revision C 2060290 201 For your notes ...
Page 51: ...Revision C CASE 3 1 2060290 201 3 Preparing the Patient ...
Page 52: ...3 2 CASE Revision C 2060290 201 For your notes ...
Page 64: ...3 14 CASE Revision C 2060290 201 Attaching the Acquisition Module For your notes ...
Page 65: ...Revision C CASE 4 1 2060290 201 4 Selecting a Patient Record and Test Type ...
Page 66: ...4 2 CASE Revision C 2060290 201 For your notes ...
Page 73: ...Revision C CASE 5 1 2060290 201 5 Resting ECG ...
Page 74: ...5 2 CASE Revision C 2060290 201 For your notes ...
Page 108: ...5 36 CASE Revision C 2060290 201 Post Test Review For your notes ...
Page 109: ...Revision C CASE 6 1 2060290 201 6 Exercise Test ...
Page 110: ...6 2 CASE Revision C 2060290 201 For your notes ...
Page 171: ...Revision C CASE 7 1 2060290 201 7 Ambulatory Blood Pressure Measurement ...
Page 172: ...7 2 CASE Revision C 2060290 201 For your notes ...
Page 188: ...7 18 CASE Revision C 2060290 201 Post Test Review For your notes ...
Page 189: ...Revision C CASE 8 1 2060290 201 8 Spirometry Test ...
Page 190: ...8 2 CASE Revision C 2060290 201 For your notes ...
Page 208: ...8 20 CASE Revision C 2060290 201 Post Test Review For your notes ...
Page 209: ...Revision C CASE 9 1 2060290 201 9 File Management ...
Page 210: ...9 2 CASE Revision C 2060290 201 For your notes ...
Page 222: ...9 14 CASE Revision C 2060290 201 Working Temporarily with the Local Database For your notes ...
Page 223: ...Revision C CASE 10 1 2060290 201 10 Operation with Administrative Tools ...
Page 224: ...10 2 CASE Revision C 2060290 201 For your notes ...
Page 227: ...Revision C CASE 11 1 2060290 201 11 DICOM Interface ...
Page 228: ...11 2 CASE Revision C 2060290 201 For your notes ...
Page 233: ...Revision C CASE 12 1 2060290 201 12 System Settings ...
Page 234: ...12 2 CASE Revision C 2060290 201 For your notes ...
Page 296: ...12 64 CASE Revision C 2060290 201 System Configuration For your notes ...
Page 297: ...Revision C CASE A 1 2060290 201 Cleaning and Maintenance A Cleaning and Maintenance ...
Page 298: ...A 2 CASE Revision C 2060290 201 Cleaning and Maintenance For your notes ...
Page 301: ...Revision C CASE B 1 2060290 201 Miscellaneous B Miscellaneous ...
Page 302: ...B 2 CASE Revision C 2060290 201 Miscellaneous For your notes ...
Page 325: ...Revision C CASE C 1 2060290 201 Order Information C Order Information ...
Page 326: ...C 2 CASE Revision C 2060290 201 Order Information For your notes ...
Page 330: ...C 6 CASE Revision C 2060290 201 Order Information Order Information ...
Page 331: ...Revision C CASE D 1 2060290 201 Technical Description D Technical Description ...
Page 332: ...D 2 CASE Revision C 2060290 201 Technical Description ...