Auxiliary O2+Air
Use the optional Auxiliary O2+Air (Aux O2+Air) switch to deliver O2
and Air through the Aux O2+Air outlet on the front of the system. No
anesthetic agent is delivered through the Aux O2+Air outlet. When
the switch is set to the Aux O2+Air position during a case, the outlet
indicator is lit and N2O flow is automatically shut off.
The Aux O2+Air outlet provides 100% O2 or a mixture of O2 and Air.
The delivery default is
100% O2
. Selecting the
O2+Air
quick key
enables both O2 and Air. Adjust the O2 and Air individually using the
flow control knobs.
Using the Aux O2+Air
1.
Set the Aux O2+Air switch to the Aux O2+Air position.
AC.24.295
2.
Select either
100% O2
or
O2+Air
on the display. See "
3.
Adjust O2 and Air individually using the flow control knobs.
4.
Set the switch to the circle circuit position to stop fresh gas flow
through the Aux O2+Air outlet.
3 Operation
2076152-001
3-39