Start a case
Use the
Start Case
menu to set the case data and start the case.
A case can be started using default settings or using custom
settings. The default settings are configured by the Super User. See
the "
" section for information on the
Start Case
menu defaults.
Default Settings selection shows the first of five default case types
when the
Start Case
menu is accessed. Four of the default case
types are configured by the Super User. The fifth default case is Last
Case.
The
Ideal Weight
,
Age
, and
Volume Apnea Alarm
values are set to
the pre-selected settings defined by the Super User corresponding to
the case type.
Make sure that the patient breathing circuit is correctly
assembled and that the ventilator settings are clinically
appropriate before starting ventilation. Incorrect breathing
circuit assembly and incorrect ventilator settings can
injure the patient.
Make sure that the preset alarm limits are appropriate for
the patient before starting ventilation. Incorrect alarm
settings can injure the patient.
In case of a patient emergency, the
Full Test
may be bypassed by
selecting
Start Case
and then selecting
Bypass
. The general
message
Please Do Checkout
is displayed if a
Full Test
or all of
the individual tests are not completed with passing results within 24
hours. The messages
Checkout has not passed.
and
Checkout
bypass will be recorded to the system log.
will be displayed in
Checkout
menu when checkout is bypassed.
Volume Apnea Alarm
is not shown on the
Start Case
menu when
the
Volume Apnea Selection
is set to
Disable
in the Super User
settings.
The
TV for Ideal Body Weight
menu item from the
Patient
Demographics
menu can only be accessed when the ventilation
mode is set to
VCV
,
PCV-VG
,
SIMV VCV
, and
SIMV PCV-VG
. Use
this setting for breath rate and tidal volume calculations based on the
set patient weight.
Minimum Alveolar Concentration
The Minimum Alveolar Concentration (MAC) concept is based on the
assumption that in a steady state condition, the alveolar partial
pressure of a gas is equal to the partial pressure in the effector organ
of the central nervous system. MAC values are used to estimate the
level of anesthesia caused by volatile anesthetics. The MAC value
display range is 0.0 to 9.9.
WARNING
•
Note
Note
Note
Carestation
™
620/650/650c (A1)
3-8
2076152-001
Summary of Contents for Carestation 620
Page 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Page 12: ...Carestation 620 650 650c A1 x 2076152 001...
Page 58: ...2 24 2076152 001...
Page 98: ...3 40 2076152 001...
Page 102: ...4 4 2076152 001...
Page 128: ...6 12 2076152 001...
Page 150: ...7 22 2076152 001...
Page 182: ...8 32 2076152 001...
Page 196: ...9 14 2076152 001...
Page 246: ...11 44 2076152 001...
Page 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...