Ventilator operating specifications
Based on the ventilator settings in use, the setting ranges may be
constrained so that the specifications stated are met.
Pneumatics
Gas source
Anesthesia system
Gas composition
Medical Air or O2
Peak gas flow
120 l/min at 240 kPa (35 psi), 0.75 seconds
Flow valve range
1 to 120 l/min at 240 kPa (35 psi)
Fresh gas compensation
Flow compensation range
0.1 to 15 l/min
Gas composition
O2, N2O, Air, anesthetic agents
Pressure
Patient airway pressure
measurement range
-20 to 120 cmH2O, +/- 1 cmH2O
resolution
High pressure alarm set range
12 to 100 cmH2O, 1 cm increment
Sustained pressure alarm range
6 to 40 cmH2O, 1 cm increment
Display range
-20 to 120 cmH2O
Volume
Tidal volume display range
less than 1 to 9999 ml, 1 ml resolution
Tidal volume setting range
20 to 1500 ml
Minute volume display range
less than 0.1 to 99.9 l/min, 0.1 liter
resolution
11 Specifications and theory of operation
2076152-001
11-37
Summary of Contents for Carestation 620