Setting
Range
Mode
VCV
PCV
PCV-VG SIMV VCV SIMV PCV SIMV
PCV-VG
PSVPro CPAP
+PSV
Rise Rate
Auto, 1-10
*
Auto
*
Auto
Auto
*
Auto
Auto
Rise Rate
1-10
*
*
5
*
*
5
*
*
PSV Rise Rate
Auto, 1-10
*
*
*
*
*
Auto
*
*
Trig Window
Off, 5-80% of
Texp
*
*
*
80
80
80
80
*
Flow Trigger
0.2-10 l/min
*
*
*
2.0
2.0
2.0
2.0
2.0
End of Breath
5-75% of
peak
inspiratory
flow
*
*
*
30
30
30
30
30
Backup Time
10-30
seconds
*
*
*
*
*
*
30
*
Exit Backup
Off, 1-5
spontaneous
breaths
*
*
*
*
*
*
2
*
Ventilation mode transition
Ventilation settings selectable though the quick keys and
More
Settings
are set to the factory default settings at start up of the
system and at the end of each case. The ventilation settings remain
at the factory default until changed by the user. If a ventilation setting
is changed, the new setting remains active in all applicable modes
until the setting is changed again. Exceptions include:
•
For modes that use RR 4 to 100, the RR transfers to the new
mode at the set RR or 60 whichever is smaller. For example, if
the mode in use has a RR of 75 and the new mode does not
support a RR of 75, the RR is set to 60.
•
For modes that use RR 2 to 60, the RR does not transfer to
modes that use RR 4 to 100. The RR is set to the factory default
or the last set if previously used during the case. For example, if
during the first case the RR (2 to 60) is set to 2 and the new
mode uses RR 4 to 100, the RR is set to the factory default.
•
For modes that use I:E, when transferring to a mode that uses
Tinsp, the Tinsp is set to a calculation of I:E or 5 seconds
whichever is lower.
•
For CPAP + PSV mode, Psupport and Pinsp settings are linked
during the patient case. When the Psupport setting is changed,
the Pinsp setting changes to the same value. The Pinsp setting
shows in the ventilation mode information above the ventilator
settings. To break the link during a case, change the Pinsp
setting using the Pinsp key.
Carestation
™
620/650/650c (A1)
11-36
2076152-001
Summary of Contents for Carestation 620
Page 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Page 12: ...Carestation 620 650 650c A1 x 2076152 001...
Page 58: ...2 24 2076152 001...
Page 98: ...3 40 2076152 001...
Page 102: ...4 4 2076152 001...
Page 128: ...6 12 2076152 001...
Page 150: ...7 22 2076152 001...
Page 182: ...8 32 2076152 001...
Page 196: ...9 14 2076152 001...
Page 246: ...11 44 2076152 001...
Page 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...