Electrical power
Supply voltage
100-120, 220-240, or 120/220-240 Vac +/- 10% at 50 or
60 Hz
Average power
consumption
Less than 55 VA
Note: Measured through the power cord when operated at the ventilation
setting VCV, TV = 500 ml/min, RR = 12/min, I:E = 1:2 with nothing plugged
into auxiliary outlets and an airway module inserted. Measurements were
taken over a 15 minute time frame after system had been running for 10
minutes.
Voltage
100-120 Vac 220-240 Vac 100-120 Vac 220-240 Vac
Outlet circuit
breakers
(2) 2 A
(1) 3 A
(2) 1 A
(1) 2 A
None
None
Mains fused
10 A
8 A
2 A
2 A
System leakage
current limit - do
not exceed
UL and CSA rated systems (U.S.A. and Canada): less
than 300 μamps for the system and all systems
connected to electrical outlets.
IEC rated systems (not U.S.A. and Canada): less than
500 µamps for the system and all systems connected to
electrical outlets.
Note: Products connected to electrical outlets may
increase the leakage current above these limits.
Resistance to
ground
Less than 0.2 Ω.
Power cord
Length
5 meters
Voltage rating
100 to 240 Vac
Current capacity
10 A for 220-240 Vac
15 A for 100-120 Vac
10 A for 120/220-240 Vac
Type
Three conductor power supply cord (medical
grade where required).
This equipment must be connected to a supply mains
with a protective earth to avoid the risk of electric shock.
WARNING
11 Specifications and theory of operation
2076152-001
11-9
Summary of Contents for Carestation 620
Page 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Page 12: ...Carestation 620 650 650c A1 x 2076152 001...
Page 58: ...2 24 2076152 001...
Page 98: ...3 40 2076152 001...
Page 102: ...4 4 2076152 001...
Page 128: ...6 12 2076152 001...
Page 150: ...7 22 2076152 001...
Page 182: ...8 32 2076152 001...
Page 196: ...9 14 2076152 001...
Page 246: ...11 44 2076152 001...
Page 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...