Pneumatic specifications
All gases supplied to the system must be medical grade.
Gas supplies
Pipeline gases
O2, Air, N2O
Cylinder gases
O2, Air, N2O (3 cylinder maximum)
Cylinder connections
Pin indexed
Nut and gland DIN-477
Large cylinder kit available for O2 and N2O
Primary regulator output
pressure
Pin indexed: The primary regulator is set to
pressure less than 345 kPa (50 psi).
DIN-477: The primary regulator is set to
pressure less than 414 kPa (60 psi).
Pressure-relief valve
Approximately 758 kPa (110 psi)
Pipeline connections
(filtered)
DISS-Male; DISS-Female; AS 4059
(Australian); S90-116 (French Air Liquide);
NIST (ISO 5359). All fittings available for O2,
Air, and N2O.
Pressure displays
On system display.
Pipeline inlet pressure
280-600 kPa (41-87 psi)
N2O shutoff
3.5 to 27.6 kPa (0.5 to 4 psi)
ACGO port relief
A relief valve on the ACGO outlet limits the fresh gas pressure at the
ACGO port to 12.25 kPa (125 cmH2O) between 25 l/min and 75 l/
min.
CAUTION