Calibration menu
Access the
Calibration
menu from the
Checkout
menu or by
selecting the
System Setup
button. Follow the instructions on the
screen.
The
Calibration
menu is not available during a case.
" section
for information on calibrating the airway modules.
Flow and pressure calibration
1.
Select
Checkout
or
System Setup
.
2.
Select
Calibration
on the
Checkout
or
System Setup
menu.
3.
Select
Flow and Pressure
.
4.
Follow the instructions on the screen.
Circuit O2 cell
Weekly O2 cell calibration
Weekly O2 cell calibration will calibrate the O2 cell to 21% O2,
through exposure to room air.
1.
Select
Checkout
or
System Setup
.
2.
Select
Calibration
on the
System Setup
menu.
3.
Select
Weekly O2 Cell
.
4.
Follow the instructions on the screen.
Monthly O2 cell calibration
Monthly O2 cell calibration will calibrate the O2 cell to 21% O2,
through exposure to room air, and to 100% O2, through exposure to
100% O2 supply.
1.
Make sure the patient connection is not plugged or there is no
patient tubing connected to the system.
2.
(ACGO option only.) Set the ACGO switch to Circle.
3.
(Aux O2+Air option only.) Set the Aux O2+Air switch to Circle.
4.
Set the Bag/Vent switch to Vent.
5.
Select
Monthly O2 Cell
.
6.
Follow the instructions on the screen.
Note
9 User maintenance
2076152-001
9-11
Summary of Contents for Carestation 620
Page 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Page 12: ...Carestation 620 650 650c A1 x 2076152 001...
Page 58: ...2 24 2076152 001...
Page 98: ...3 40 2076152 001...
Page 102: ...4 4 2076152 001...
Page 128: ...6 12 2076152 001...
Page 150: ...7 22 2076152 001...
Page 182: ...8 32 2076152 001...
Page 196: ...9 14 2076152 001...
Page 246: ...11 44 2076152 001...
Page 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...