Parameters setup
Use the
Spirometry
menu to change the monitoring settings of the
data source, CO2, O2, agent, and spirometry. To access the
Spirometry
menu, push the
Spirometry
button and then select
Setup Loops
.
Data source
Several monitoring parameters can be obtained from the ventilator or
the airway module. Information retrieved from the airway module is
identified with the module data indicator.
Figure 6-5 • Airway module data indicator
Set the
Data Source
to
Patient
or
Vent
to select the primary source
for information. If
Patient
is selected, the airway module will be the
first source for information. If
Vent
is selected, the internal sensors of
the ventilator will be the first source for information.
When using the Airway Gas Option, only
Vent
is available for the
Data Source
.
When setting the airway module as the data source, make sure that
a D-lite
™
or Pedi-lite sensor is properly connected on the airway
module. If the sensor is not properly connected, but the airway
module is installed, the waveform shows no flow. The gas monitor
samples and displays room air.
If information is not available through the airway module, information
comes from the internal ventilator sensors.
Carestation
™
620/650/650c (A1)
6-8
2076152-001
Summary of Contents for Carestation 620
Page 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Page 12: ...Carestation 620 650 650c A1 x 2076152 001...
Page 58: ...2 24 2076152 001...
Page 98: ...3 40 2076152 001...
Page 102: ...4 4 2076152 001...
Page 128: ...6 12 2076152 001...
Page 150: ...7 22 2076152 001...
Page 182: ...8 32 2076152 001...
Page 196: ...9 14 2076152 001...
Page 246: ...11 44 2076152 001...
Page 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...