Full test
The
Full Test
or the individual tests must be performed at least once
within every 24-hour period.
Perform the
Full Test
at the start of each day. The full test runs
automatically and the color of the test indicator changes when the
test is finished or if user action is required.
Perform a
Full Test
when any component of the system is changed
(breathing system, vaporizers, pipeline inlets).
The
Full Test
does the following tests:
Ventilator Leak
,
Vaporizer
Leak
,
Gas Controls
, and
Circuit leak
. When one of the tests is
completed, the next test begins.
1.
From the
Checkout
menu, select
Full Test
and follow the
instructions.
2.
If a test fails, follow the instructions to perform a retest or accept
the results.
3.
Before the test results are shown, the following checklist is
shown for confirmation:
•
‘Verify O2 flush operates correctly.’
•
‘Connect respiratory gas monitor to the patient circuit.’
•
‘Check that patient monitor settings are correct.’
•
‘Check that backup ventilation is available and operational.’
•
‘Check that patient suction is set up correctly.’
•
‘Check the pressure of the backup O2 supply.’
4.
When the
Full Test
is completed, start a case.
Make sure that a backup ventilation method (not integrated into the
system) is available and operational. For example, a bag and mask
respirator.
Contact a trained service representative to add checklist items. Four
checklist items can be added to match unique facility procedures.
In case of a patient emergency, the
Full Test
may be bypassed by
selecting
Start Case
and then selecting
Bypass
. The general
message
Please Do Checkout
is displayed if a
Full Test
or all of
the individual tests are not completed with passing results within 24
hours. The messages
Checkout has not passed.
and
Checkout
bypass will be recorded to the system log.
will be displayed in
Checkout
menu when checkout is bypassed.
The general message
Redo Gas Controls Checkout
shows if the
N2O is reconnected to the system after a
Full Test
was done without
the N2O connected. Perform a
Gas Controls
test to remove the
message.
Note
Note
Note
Note
5 Preoperative tests
2076152-001
5-7
Summary of Contents for Carestation 620
Page 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Page 12: ...Carestation 620 650 650c A1 x 2076152 001...
Page 58: ...2 24 2076152 001...
Page 98: ...3 40 2076152 001...
Page 102: ...4 4 2076152 001...
Page 128: ...6 12 2076152 001...
Page 150: ...7 22 2076152 001...
Page 182: ...8 32 2076152 001...
Page 196: ...9 14 2076152 001...
Page 246: ...11 44 2076152 001...
Page 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...