EEG and auditory evoked potentials
Problem
Solution
●
Check the number of channels.
The screen configuration may not include the EEG
parameter window.
●
Check the monitor setup.
Why are the numeric values replaced by ---?
The patient has high muscle activity in the head area,
or noise from some interfering equipment is coupling
to electrode cables.
●
Relax the patient and remove the source of noise.
Why do the electrode impedances show --- and
there is a message prompting to check the ground
electrode?
The ground electrode is poorly connected to the
patient, or its cable is not connected to the headbox.
●
Check the electrode and cable.
If the impedance of the electrode is too high, the
measurement fails even if the electrode is properly
attached.
●
Use better electrodes or prepare the skin better.
Why are there some baseline fluctuations?
For instance, sweating may cause variations in the
electrode impedance.
●
In case the fluctuation is disturbing, prepare the
skin and replace the electrode.
Measuring auditory evoked potentials (AEP)
Preparing the patient for AEP measurement
1.
Prepare the patient and electrodes as you would for EEG measurement.
2.
Connect the leadset and electrodes.
3.
Connect earphones to the headbox and place them on the patient's ear/s.
NOTE
Make sure that the earphones do not press the patient's
head in any position (for example, when the patient is
lying on one side).
NOTE
Use only GE Healthcare supplied headphones/earphones.
In other headphones/earphones the sensitivity may vary
causing too high an intensity, which may damage the ear,
or too low an intensity, which may cause a poor response.
2062971-001
CARESCAPE Modular Monitors
387
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...