Patient Spirometry
Patient Spirometry troubleshooting
Problem
Solution
Values seem erroneous
●
Check the patient status.
●
Check that you are using the correct sensor type:
D-lite(+) for adult patients, Pedi-lite(+) for pediatric
patients.
●
Check the sensor type selection.
●
Check that the Patient Spirometry tube connectors
and their connections are tight and not leaking.
●
Check the arterial blood gas values.
●
Check that the sampling line is not kinked.
Values seem unstable
●
Remove the D-lite(+)/Pedi-lite(+) and shake drops
away.
●
Check that the connectors on the
D-lite(+)/Pedi-lite(+) are intact and that
connections are tight.
Strong vibrations in the loop
●
Check the patient status.
●
Check the patient and system for water or
secretions.
Why do the monitor and the ventilator show a
different compliance value?
The monitor measures directly at the endotracheal
tube and reflects more accurate compliance of
the lung. Hoses and other equipment between
the patient and the ventilator influence the values
measured in the ventilator.
2062971-001
CARESCAPE Modular Monitors
337
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...