Mixed venous oxygen saturation (SvO₂)
●
An in vitro calibration was completed with this optical module using the same
catheter within the last 24 hours.
1.
Select the SvO
2
parameter window.
2.
Select the
Calibration
tab.
3.
Select
Recall Calibration
.
Calibrating SvO
2
in vivo
NOTE
For optimal accuracy, perform in vivo calibration at least
every 24 hours.
Follow the
Advice to User
information displayed in the
Calibration
window to guide
you through the calibration steps.
1.
Select the SvO
2
parameter window.
2.
Select the
Calibration
tab.
3.
Select
In Vivo Calibration
>
Calibrate
to perform in vivo calibration using the
SvO
2
and Hb values measured from the blood sample.
4.
Select
Draw Blood Sample
and slowly draw a blood sample.
5.
Enter the laboratory results:
a. Select a value for
Lab ScvO2
.
b. Select a value for
Lab Hb
.
6.
Select
Save Lab Values
to complete calibration.
Updating the Hb value for SvO
2
measurement
The patient’s Hb value should be updated at least every 24 hours as it affects the
SvO
2
measurement. The
Update Hb
selection is active after in vivo calibration has
been completed with the
Save Lab Values
.
1.
Select the SvO
2
parameter window.
2.
Select the
Calibration
tab.
3.
Select
Update Hb
and set the value.
Setting the SvO
2
alarms
1.
Select the SvO
2
parameter window.
2.
Select the
Alarms
tab.
If alarms have been set to
Alarm Off
, the alarm limits are greyed out. Select
Alarm On
to set the alarms.
3.
Set the alarm limits.
Stopping the SvO
2
measurement
1.
Remove the catheter from the patient.
2.
Disconnect the catheter from the optical module.
3.
Disconnect the optical sensor from the SvO
2
module.
4.
Discard the catheter.
278
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...