Cardiac output
catheters which show up in the list can be added or deleted through
Care Unit
Settings
>
Parameters
>
Catheters
. These settings are password protected.
For more information, see the CARESCAPE Modular Monitors Supplemental
Information Manual.
Selecting a C.O. catheter from the list
1.
Select the cardiac output parameter window.
2.
Select the
Setup
tab.
3.
Select a catheter name from the
Manufacturer
list.
4.
Select a catheter model from the
Model
list.
Entering a user-defined C.O. catheter
All user-defined catheter settings are erased when the monitor is discharged.
1.
Select the cardiac output parameter window.
2.
Select the
Setup
tab.
3.
Select
User Defined
from the
Manufacturer
list.
4.
Set the
Injectate Volume
to match the value listed on the catheter packaging.
5.
Set the
Computation Constant
to match the value listed on the catheter
packaging:
●
Ice Cold
: temperature below +6°C.
●
Room Temp
: temperate above +18°C.
●
Any other injectate temperature value b6°C and +18°C.
Selecting the C.O. injectate probe type
NOTE
PDM and TRAM only. E-modules detect the type of injectate
probe automatically.
1.
Select the cardiac output parameter window.
2.
Select the
Setup
tab.
3.
Select the correct
Probe Type
:
Bath
or
In-Line
.
Setting a C.O. right ventricular ejection fraction (REF)
measurement
NOTE
E-COP and E-COPSv modules and catheters that support right
ventricular ejection fraction measurement only. PDM and
TRAM do not provide a REF measurement.
A valid heart rate is required to take a REF measurement. ECG from a telemetry
transmitter cannot be used.
1.
Select the cardiac output parameter window.
2.
Select the
Setup
tab.
3.
Select the check box for
REF Measurement
.
2062971-001
CARESCAPE Modular Monitors
267
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...