Invasive pressures
4.
You can save these cursors by selecting
Save
.
This will restart the waveforms. If you do not want to save the cursors, select
Restart Waveforms
.
NOTE
You can also use the
Optimize Scale
function in the invasive
pressure
Setup
menu to set the scale for manual SVP
measurement.
PA catheter insertion
The catheter insertion mode optimizes and enlarges the PA waveform field during
SWAN-GANZ thermodilution catheter insertion. Waveforms display at a rate of 12.5
mm/s and appear in the following display order:
ECG1
,
Art
,
CVP
,
PA
.
The arterial priority order is:
Art
,
ABP
,
Fem
, or
UAC
.
Selecting the PA catheter insertion mode
1.
Select the PA invasive pressure parameter window.
2.
Select
Zero
to zero the invasive pressure channel.
3.
Select
Catheter Insertion
.
The pressure scale settings in the
Catheter Insertion
window follow the scale
settings in the
Setup
menu.
4.
Select the procedure:
●
To start an SvO
2
procedure, select
SvO2
.
●
To start a pulmonary capillary wedge pressure procedure, select
Wedge
.
●
To start a cardiac output procedure, select
C.O.
In the
Catheter Insertion
menu, you can also freeze or restart the waveforms, and
print them:
●
To freeze the moving waveforms, select
Freeze Waveforms
. At any time, select
Restart Waveforms
to restart the waveforms.
●
To print the catheter insertion waveforms, select
Print Waveforms
. At any time,
select
Stop Printing
or
Cancel Printing
to stop printing the waveforms.
Pulmonary capillary wedge pressure (PCWP)
measurement
You can obtain a PA wedge measurement (PCWP) manually or with the automated
wedge program. The manual measurement mode allows you to manually determine
the PCWP value. The automated wedge program displays on-screen messages to
inflate or deflate the catheter balloon. In either mode, the wedge algorithm then
determines the PCWP value. You can confirm this value or adjust the measurement
with the provided cursor.
NOTE
E-modules always use the automated wedge program to
determine the PCWP value.
246
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...