Invasive pressures
3.
Set the alarm limits.
Invasive pressures alarm priorities
You can select priorities for the
Art high/low
,
ABP high/low
,
Fem high/low
,
FemV
high/low
,
CVP high/low
,
PA high/low
,
RAP high/low
,
RVP high/low
,
LAP high/low
,
ICP high/low
,
CPP high/low
, and
P1 high/low
to
P8 high/low
alarms through
Alarm
Setup
>
Alarm Priorities
>
Invasive Pressures
. The choices are
Low
,
Medium
,
High
,
and
Escalating
.
Systolic pressure variation and pulse pressure
variation
Systolic pressure variation (SPV) and pulse pressure variation (PPV) can provide useful
information for example when assessing the effects of fluid therapy on the cardiac
output of a patient. A parameter window with both SPV and PPV values appears on
the display provided that SPV has been selected to the screen and the arterial site
selected as the SPV source is active.
The SPV and PPV measurement is automatic, and SPV can also be taken manually.
With the NICU software package, no automatic SPV or PPV are available , only the
manual SPV can be used.
NOTE
The SPV and PPV measurements are reliable for mechanically
ventilated patients with no arrhythmias, and when the arterial
site selected as the SPV source is providing reliable readings.
Changing the SPV source
1.
Select the invasive pressure parameter window.
2.
Select
Art
,
ABP
,
Fem
or
UAC
.
3.
Select the
Setup
tab.
4.
Select
SPV Source
>
Art
,
ABP
,
Fem
, or
UAC
. You can also turn off the measurement
by selecting
Off
(default).
Measuring SPV manually
The SPV can also be measured manually. In addition to ECG1 and the selected SPV
source, one of the following parameters is displayed in this order: Paw, CO
2
, Resp. You
can set the SPV cursors to define the difference between the minimum and maximum
systolic peak pressures.
1.
Select the SPV and PPV parameter window.
2.
Select
Freeze Waveforms
.
3.
Adjust the cursors with the arrow selectors.
NOTE
You must always adjust the cursors in the NICU software
package. The monitor does not suggest any cursor
positions as there is no automatic SPV measurement in
this software package.
2062971-001
CARESCAPE Modular Monitors
245
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...