Invasive pressures
2.
Select the
Setup
tab.
3.
Select
Show CPP
.
Selecting Smart BP
NOTE
PDM, TRAM and Tram-Rac modules only.
Art
,
ABP
, and
Fem
invasive pressure channels only.
Smart BP
is an algorithm that temporarily deactivates the arterial and femoral alarms
when it detects the zeroing of a transducer, fast flushing of the system, or blood
draws. The message
Artifact
displays during the alarm deactivation. When pulsatile
pressure returns and 15-20 beats are detected, numerics are displayed and alarms
are reactivated.
1.
Select the invasive pressure parameter window.
2.
Select the
Advanced
tab.
3.
Select
Smart BP
.
Compensating for intra-aortic balloon pump (IABP)
waveform irregularities
WARNING
— INCORRECT PULSE RATE — Be sure to turn off the IABP
setting when the cardiac assist device is no longer used.
Failure to do so could result in incorrect pulse rate readings.
CAUTION
— PATIENT HAZARD — If you choose to trigger the balloon
pump from the monitor, contact the balloon pump
manufacturer directly for interface requirements, as they
vary among manufacturers. Some trigger modes on certain
balloon pump devices may not be compatible with the GE
analog output signal, and use may contribute to patient injury
or sub-optimal pumping results.
NOTE
PDM, TRAM, and Tram-Rac modules only.
Art
,
ABP
, and
Fem
invasive pressure channels only.
1.
Select the invasive pressure parameter window.
2.
Select the
Advanced
tab.
3.
Select
IABP On
.
IABP
now displays in the invasive pressure channel parameter window.
Selecting the invasive pressure response time
NOTE
E-modules only.
1.
Select the invasive pressure parameter window.
2.
Select the
Setup
tab.
3.
Select
Normal
or
B-to-B
(beat-to-beat) from the
Response
list.
2062971-001
CARESCAPE Modular Monitors
243
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...