Invasive pressures
NOTE
This setting adjusts the primary heart rate source for all of the
hemodynamic parameters.
NOTE
This setting is available for
Art
,
ABP
,
Fem
, or
UAC
invasive
pressure channels only.
UAC
is only available in the NICU
software package.
NOTE
HR Alarms
must be configured as
Single
to enable invasive
pressure as the primary heart rate source.
1.
Select the invasive pressure parameter window.
2.
Select the
Setup
tab.
3.
Select the heart rate source from the
Primary HR Source
list.
Variable beat tone
You can configure a variable beat tone through
Monitor Setup
>
Default Setup
>
Care
Unit Settings
>
Parameters
>
Variable Beat Tone
. This setting is password protected.
If it set to
All beat sources
, the SpO
2
saturation affects all beep sounds including ECG
and IP when the SpO
2
measurement is available: beep frequency changes according
to increasing and decreasing SpO
2
values. If the setting is set to
Only SpO2
, other beep
sounds are not affected by the changing SpO
2
values.
For more information, see the CARESCAPE Modular Monitors Supplemental
Information Manual.
Selecting the ventilation mode
NOTE
E-modules only.
This setting affects the respiration filter.
1.
Select the invasive pressure parameter window.
2.
Select the label (
P2
,
P5
,
P6
,
P7
,
P8
,
CVP
,
FemV
,
PA
,
RAP
,
RVP
,
LAP
).
3.
Select
Ventilation Mode
>
Spontaneous
or
Controlled
.
Showing the pulse rate in the invasive pressure parameter
window
NOTE
This setting is available for
Art
,
ABP
,
Fem
, or
UAC
invasive
pressure channels only.
1.
Select the invasive pressure parameter window.
2.
Select the
Setup
tab.
3.
Select
Show Pulse Rate
.
Showing the CPP value in the ICP parameter window
A valid mean arterial pressure is required to compute the cerebral perfusion pressure
(CPP) value.
1.
Select the invasive pressure parameter window.
242
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...