Invasive pressures
Selecting the display mode for IP waveforms
You can select the invasive pressure waveforms to be shown as individual waveforms,
or in a combined view.
1.
Select
Monitor Setup
>
Screen Setup
.
2.
B850 and B450 with the Double Video license: Select
Screen 1
or
Screen 2
tab.
3.
Select
Upper Parameter Area
.
4.
Select an option from the
Invasive Pressure Waveforms
list:
●
To view individual waveforms, select
Individual
.
●
To combine the currently displayed adjacent waveforms (2 to 4), select
Combined
. The new waveform field will use the combined height of the original
fields.
●
To combine up to four waveforms in one field, select
4invP
. The new waveform
field will use the height of two upper parameter windows.
Using the invasive pressure measurement
Invasive pressure measurement mapping
Invasive pressure measurements are mapped to one of eight (B850), seven (B650), or
six (B450) invasive pressure channels as follows:
Pressure channel
Pressure measurement source
P1
PDM (Masimo), PDM (Nellcor), TRAM 451, TRAM 451N5, TRAM
451M, E-PSMP.
P2
PDM (Masimo), PDM (Nellcor), TRAM 451, TRAM 451N5, TRAM
451M, E-PSMP.
P3
PDM (Masimo), PDM (Nellcor), TRAM 451, TRAM 451N5, TRAM
451M, .
P4
PDM (Masimo), PDM (Nellcor), TRAM 451, TRAM 451N5, TRAM
451M, .
P5
Dual BP, BP/CO, or BP/Dual Temp Tram-Rac module in Tram-Rac
slot 3, or E-PP.
P6
Dual BP or BP Tram-Rac module in Tram-Rac slot 3, E-PP.
P7
Dual BP, BP/CO, or BP/Dual Temp Tram-Rac module in Tram-Rac
slot 4, E-P, or E-PT.
P8
Dual BP or BP Tram-Rac module in Tram-Rac slot 4, E-COP, or
E-COPSv.
238
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...