Invasive pressures
●
Do not turn on the IABP algorithm unless a balloon pump is in use.
●
If two pressure measurement modules that map to the same pressure channel
are connected, the first detected pressure measurement module is assigned to
the indicated pressure channel.
●
A pressure channel is activated when a pressure transducer interface cable is
connected to the PDM (Masimo), PDM (Nellcor), Tram 451, Tram 451N5, Tram
451M, Dual BP, BP/CO, BP, or BP/Dual Temp, or E-COP module, or when a pressure
transducer is connected to the E-COPSv, E-PSMP, E-PP, E-P, or E-PT module.
●
A pressure channel is deactivated when the pressure transducer interface cable is
disconnected from the PDM (Masimo), PDM (Nellcor), Tram 451, Tram 451N5, Tram
451M, Dual BP, BP/CO, BP, or BP/Dual Temp, or E-COP module, or when the pressure
transducer is disconnected from the or E-COPSv, E-PSMP, E-PP, E-P, or E-PT module.
●
E-modules: A deactivated pressure channel does not release the assigned pressure
channel. To release an assigned pressure channel, remove the module from the
system.
●
PCWP: Follow your care unit’s policy and procedures for obtaining PCWP
measurements, including balloon inflation duration.
●
SPV and PPV: Measurement is reliable only when the patient is mechanically
ventilated.
Invasive pressure measurement setup
Invasive pressure equipment to patient connection
1.
Module with invasive pressure measurement capability
2.
Fluid bag with pressure infusor
3.
Transducer setup
4.
Invasive blood pressure adapter cable; single or dual cable (optional)
Invasive pressure module keys
There are invasive pressure keys on the following modules:
236
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...