Non-invasive blood pressure
1.
Cuff deflation
2.
Systolic pressure (ratio of maximum amplitude)
3.
Mean arterial pressure (maximum pulsation amplitude)
4.
Diastolic pressure (ratio of maximum amplitude)
5.
Cuff pulsations (each pulsation represents one heart beat)
6.
Amplitude (changes based on cuff pressure)
7.
Resulting waveform
When the measurement signal detects artifacts, the deflation is stopped until a good
signal can be measured again. The cuff pressure remains at one pressure level until it
gets two similar pulses before continuing to the next pressure level. It compares two
sequential pulses and uses information and statistics from previous pulses.
The performance in disturbance or artifact cases is comparable to DINAMAP Classic.
The measurements do not use ECG information. The initial inflation pressure cannot
be adjusted, it is determined by the patient type (adult, child or infant). The first
determination initially pumps up to a target cuff pressure of 170 mmHg for adult, 150
mmHg for child or 120 mmHg for infant. After the first measurement, the inflation
pressure is adjusted according to measured systolic pressure.
NIBP calibration
NIBP calibration procedure is explained in the Module Frames and Modules Technical
Manual. The calibration procedure is password protected.
232
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...