Non-invasive blood pressure
2.
Stop the measurement by pressing the
Start Cancel
key again.
Automatic NIBP measurements
Automatic NIBP measurements and monitor clock
synchronization
NOTE
TRAM modules do not support automatic clock
synchronization.
Clock synchronization timing (cycling sync) automatically synchronizes your automatic
NIBP measurement time intervals with the monitor clock. For example, if automatic
measurements are initiated for five minute intervals at 4:02, the first measurement is
taken immediately at 4:02. The next measurement will be taken at 4:05 (interval and
clock are now synchronized). All measurements will continue at five minute intervals
(i.e., 4:10, 4:15, etc.)
There will always be at least a 30 second delay between two consecutive NIBP
measurements during auto cycling. If an automatic measurement completes with
less than 30 seconds to the next scheduled measurement, the monitor will delay the
scheduled measurement until 30 seconds have passed. The cycling synchronization
is not done during these 30 seconds but it will be done after the delayed auto
measurement starts.
Examples with a 5 minute cycling time:
Completion with less than 30 seconds to
the next scheduled measurement
Completion after the next scheduled
measurement should have started
●
First auto measurement starts: 4:59:00
●
First auto measurement completes:
4:59:40
●
Second auto measurement starts: 5:00:20
(not clock synchronized)
●
Third auto measurement starts: 5:05:00
(clock synchronized)
●
First auto measurement starts: 4:59:00
●
First auto measurement completes:
5:00:10
●
Second auto measurement starts: 5:05:00
Setting the cycle time between NIBP measurements
To automatically measure NIBP at set time intervals, you must first set the cycle time.
1.
Select the NIBP parameter window.
2.
Select the cycle time from the
Cycle Time
list.
NIBP Auto mode
The
NIBP Auto
mode initiates repeated measurements for the selected
Cycle Time
.
There will be at least a 30 second delay between two consecutive NIBP measurements
during auto cycling.
Starting or stopping the
NIBP Auto
from the
NIBP Setup
menu
1.
Select the NIBP parameter window.
2.
Select
Start Cycling
for
NIBP Auto
.
224
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...