Non-invasive blood pressure
2.
For children and when using hoses without identification with a PSM, the inflation
limit must be set manually.
3.
PDM and TRAM: Always select the NIBP cuff size before starting a measurement. If
you are trying to start the measurement without selecting the cuff size first, the
NIBP
menu opens automatically with the
Cuff Size
list open.
4.
PSM: Check that the cuff hose was detected (if cuff detection is supported).
5.
PSM: Start the Venous Stasis mode and check that the pump is not restarting
during the measurement. If it does, the cuff may be leaking.
NIBP measurement on screen
When a
STAT
or
Venous Stasis
mode is enabled, a time progress bar displays in the
NIBP parameter window:
With
NIBP Auto
mode this may be replaced by a count down indicator. The selection
of either the
Graphical
progress bar or the
Numerical
indicator is a care unit setting
and it is password protected.
●
NIBP Auto
: The length of the green bars indicates the proportion of the cycling time
since the last measurement while the distance between the end of the green bar
and the last tick mark indicates the time remaining until the next measurement.
●
STAT
or
Venous Stasis
: The bar indicates the time that the mode will continue.
STAT
holds multiple measurements for 5 minutes, and
Venous Stasis
continues without
measurements for 1 minute with infants and 2 minutes with children and adults.
●
During clock synchronization, the length of the green bars indicates the difference
between the cycling time and time remaining until the next measurement.
Manual NIBP measurements
Starting or stopping a single NIBP measurement from the
main menu
1.
Start the measurement by selecting
NIBP Start
.
2.
Stop the measurement by selecting
NIBP Cancel
.
Starting or stopping a single NIBP measurement from the
NIBP Setup
menu
1.
Select the NIBP parameter window.
2.
Start the measurement by selecting
Start Manual NIBP
.
3.
Stop the measurement by selecting
Cancel NIBP
.
Starting or stopping a single NIBP measurement with the
PSM module key
1.
Start the measurement by pressing the
Start Cancel
key.
2062971-001
CARESCAPE Modular Monitors
223
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...