Impedance respiration
Lead
Description
Available with modules
Lead I
Best for detecting thoracic
breathing, but is more
susceptible to cardiogenic
artifact.
TRAM, PDM, PSM
Lead II
Equally good at detecting
thoracic or abdominal
breathing, but is more
susceptible to cardiogenic
and motion (head, neck, or
arm) artifact.
TRAM, PDM, PSM
PSM uses the ECG lead selection if the
measurement mode is 3-lead and Lead
II if the measurement mode is 5-, 6-,
or 10-leadwire set. If any electrode is
disconnected, the remaining lead (I, II or
III) is used, if possible.
Lead RL-LL
Best at detecting abdominal
breathing and is not as
susceptible to cardiogenic or
motion artifact.
The RL-LL respiration lead is
available with PDM only.
PDM
Not available for 3-lead measurement.
If you are monitoring with a fixed-lead, 3-lead cable, respiration can only be obtained
from the lead for which the cable is manufactured. For example, if the cable is a
fixed lead II cable, as indicated by a label on the cable itself, respiration can only
be obtained from lead II.
PDM and TRAM: Even though the same electrodes are used for ECG and respiration
monitoring, it is possible to get a lead fail message for respiration without one for
ECG. The impedance may be too high for respiration detection, but the electrode is
still good for ECG.
Respiration lead I electrode placement
ECG lead I for upper chest breather
IEC
AAMI/AHA
Electrode placement
R (red)
RA (white)
Just below the right clavicle.
L (yellow)
LA (black)
Just below the left clavicle.
194
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...