ECG
About the reference QRS
TRAM modules save the initial reference QRS after 64 beats have been detected. With
TRAM one reference QRS complex can be stored manually.
PDM and E-modules save the initial reference QRS up to three minutes after the ECG
measurement is started. With PDM or E-modules up to six additional reference QRS
complexes can be stored manually.
Each QRS reference is identified with the date and timestamp.
If a new reference QRS complex is saved and there is no room for another reference
complex, then the oldest manual or automatic reference complex is erased.
Saving a reference QRS manually
This selection is not available with combination monitoring.
You cannot save a reference QRS manually until an initial reference QRS complex has
been saved, or if ST Analysis is disabled.
1.
Select the ST parameter window.
2.
Select
Realtime View
.
3.
Select
Save Reference
.
The current QRS becomes the new reference QRS.
Automatic saving of reference QRS complexes
The QRS reference is not saved during an ST alarm condition, only the ST snapshot
is saved.
If a new reference QRS complex is saved and there is no room for another complex,
then the oldest manual or automatic complex is erased. You may want to manually
erase QRS complexes to avoid automated erasing.
A reference QRS complex is saved automatically whenever you do one of the following:
●
Change the Va or Vb lead.
●
Change the ST point manually.
Selecting a saved reference QRS complex for display
This selection is not available with combination monitoring or TRAM.
You can select and display a saved reference QRS for ST analysis.
1.
Select the ST parameter window.
2.
Select
Realtime View
.
3.
Select a saved reference from the
Reference QRS
list.
Erasing a reference QRS
This selection is not available with combination monitoring or TRAM.
You cannot erase the initial reference QRS.
1.
Select the ST parameter window.
2.
Select
Realtime View
.
2062971-001
CARESCAPE Modular Monitors
185
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...