ECG
●
E-modules and PDM automatically set the ST point according to the heart rate.
Manual adjustments may be required if the following automatic settings are not
adequate for example when QT time is short:
■
If the heart rate is greater than or equal to 120 bpm, then the ST point is set
to J + 60 ms.
■
If the heart rate is less than 120 bpm, then the ST point is set to J + 80 ms.
●
TRAM modules set the ST point automatically according to the profile default.
Manual adjustments may be required if the default is not adequate.
Manually adjusting the
ST Point
,
ISO Point
, or
J Point
overrides the automatic
detection of the ST point. As a result, you are responsible for monitoring the patient
ST levels with new adjustments and required to make further setting adjustments as
necessary according to changes in the patient’s rhythm.
1.
Select the ST parameter window.
2.
Select
Setup
.
3.
Select a value from the
ST Point
list.
Adjusting the isoelectric measurement (ISO) point
This selection is not available with TRAM or combination monitoring.
E-modules and PDM automatically set the isoelectric point. Manual adjustments may
be required if, for example, a P-wave is attached to the QRS-wave.
1.
Select the ST parameter window.
2.
Select
Setup
.
3.
Adjust the
ISO Point
with the arrows.
When the
ISO Point
is adjusted, also the ST point changes accordingly and its
automatic setting is stopped.
Adjusting the J point
This selection is not available with TRAM or combination monitoring.
1.
Select the ST parameter window.
2.
Select
Setup
.
3.
Adjust the
J Point
with the arrows.
When the
J Point
is adjusted, also the ST point changes accordingly.
About the realtime QRS/ST complexes
This feature is not available with TRAM or combination monitoring.
The initial reference QRS is stored and used as a comparison against the incoming
QRS complexes. The current complex is superimposed over the reference complex in
order to visually assess the change in each QRS complex. QRS complexes are updated
every 40 seconds with PDM and every 10 seconds with E-modules. Current ST values
in millimeters shall be displayed in each QRS window. Numeric values are updated
every 10 seconds.
184
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...