ECG
4.
Select a value from the
Pause Interval
list.
Setting the SVT length
E-modules and PDM only.
This setting determines how many consecutive SVCs are needed to trigger the
SV
Tachy
alarm.
1.
Select the HR parameter window.
2.
Select the
Arrhythmia
tab.
3.
Select
Atrial Alarms
.
4.
Select a value from the
SVT Length
list.
Setting HR for SVT
E-modules and PDM only.
This setting determines the minimum value for the HR to trigger the
SV Tachy
alarm.
1.
Select the HR parameter window.
2.
Select the
Arrhythmia
tab.
3.
Select
Atrial Alarms
.
4.
Select a value from the
HR for SVT /min
list.
Arrhythmia alarm messages
NOTE
A clinician must analyze the arrhythmia information in
conjunction with the other clinical findings.
Alarm message
Arrhythmia analysis
Alarm priority default
Arrhythmia detection
criteria
A Fib
Full
(All other software
packages except NICU;
not available with TRAM)
According to priority
setting
Absence of P-waves and
irregular RR-interval.
Accel. Ventric.
Full
According to priority
setting
Accelerated ventricular
rhythm - Run of PVCs
with a run length of at
least six beats and the
rate requirements have
not met for
V Tach
or
V
Brady
.
Asystole
Lethal
High
HR decreased to zero.
Bigeminy
Full
According to priority
setting
Every other beat is PVC
(N-V-N-V-N-V ).
Brady
Lethal
with NICU
software package
Full
with all other
software packages
TRAM and combination
monitoring only
According to priority
setting
Displayed 4/8-beat
average ECG heart
rate falls below the
user-selected common
HR low limit or ECG HR
low limit.
178
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...