ECG
Shown on the monitor
Shown at remote devices like central
stations
Multifocal PVCs
PVC
Missing beat
Pause
Setting the arrhythmia category to alarm
Depending on what has been allowed in the
Care Unit Settings
>
Parameters
>
ECG
>
Allowed Arrh. Levels
, you can select different arrhythmia categories to alarm.
1.
Select the HR parameter window.
2.
Select the
Arrhythmia
tab.
3.
Select
Lethal Alarms
.
4.
Select the arrhythmia category you want to alarm:
●
Full
: All arrhythmias alarm.
●
Lethal
: Only lethal arrhythmias alarm.
●
Off
: No arrhythmia alarms are generated.
Setting arrhythmia alarms
While monitoring ECG, you can adjust the settings for arrhythmia alarm conditions.
1.
Select the HR parameter window.
2.
Select the
Arrhythmia
tab.
3.
Select
Lethal Alarms
,
Ventricular Alarms
, or
Atrial Alarms
.
Ventricular Alarms
and
Atrial Alarms
are only adjustable with the full arrhythmia
license.
4.
Select an arrhythmia from the list.
5.
Select the arrhythmia alarm’s
Alarm Priority
with the arrows.
6.
Select the check box for
Create Snapshot
if you wish to activate an arrhythmia
snapshot creation.
7.
Select the check box for
Print on Alarm
if you wish to activate printing during
arrhythmia alarm.
For arrhythmia alarm waveform printing, the printing will continue until 20
seconds has passed from the clearance of the last active arrhythmia alarm (e.g.,
10 seconds saved data, arrhythmia alarm du 20 seconds data).
Setting the alarm pause interval
With TRAM or combination monitoring this setting is always 3 seconds and cannot
be edited.
You can set the time interval between the two adjacent beats before the pause alarm
condition is annunciated.
1.
Select the HR parameter window.
2.
Select the
Arrhythmia
tab.
3.
Select
Atrial Alarms
.
2062971-001
CARESCAPE Modular Monitors
177
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...