ECG
Showing ST in the HR parameter window
This option is available with the Multi-lead ST Analysis license only.
1.
Select the HR parameter window.
2.
Select the
Advanced
tab.
3.
Select the
Show ST
check box to display ST in the HR parameter window.
Showing PVC in the HR parameter window
This option is enabled with the full arrhythmia license and the
Full
detection level only.
1.
Select the HR parameter window.
2.
Select the
Advanced
tab.
3.
Select the
Show PVC
check box to display PVC in the HR parameter window.
Showing QT in the HR parameter window
This option is available with the Multi-lead QT/QTc Analysis license only.
This selection is not available with combination monitoring.
1.
Select the HR parameter window.
2.
Select the
Advanced
tab.
3.
Select the
Show QT
check box to display QT in the HR parameter window.
Displaying the ECG grid
You can have a reference grid in the
ECG1
,
ECG2
, and
ECG3
waveform areas. The grid
points will be at 200 ms horizontally and 0.5 mV vertically.
1.
Select the HR parameter window.
2.
Select the
Advanced
tab.
3.
Select the
ECG Grid
check box to display the grid.
ECG alarm limits
The
HR Alarms
can be set to
Single
or
Multiple
through
Monitor Setup
>
Default
Setup
>
Care Unit Settings
>
Parameters
>
ECG
. This setting is password protected.
The
Single
heart rate setting allows you to set one common HR limit for multiple
sources (e.g., ECG, SpO
2
, Art) and the PVC and SVC alarm limits for ECG from the
Alarms
tab. With this setting activated, turning off the SpO
2
HR alarm limits also turns
off the primary HR alarm and adjusting the SpO
2
HR limit values also adjusts the
primary HR limit value.
The
Multiple
heart rate setting allows you to set a primary heart rate/pulse rate source
and up to six individual heart rate/pulse rate alarms and limits from the
HR/PR Alarms
tab. It also allows you to set PVC and SVC alarm limits for ECG from the
PVC/SVC
Alarms
tab. (The tab is called
PVC Alarms
with TRAM or telemetry transmitters). With
the
Multiple
heart rate setting activated, turning off the SpO
2
HR alarm limits does not
turn off the primary HR alarm.
2062971-001
CARESCAPE Modular Monitors
163
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...