ECG
NOTE
Patient’s age affects the alarm limits and alarm priorities
and also the configuration of the ECG algorithm including
arrhythmia alarms in the combination monitoring mode.
For more information, see the ApexPro Telemetry System
Operator’s Manual.
When combination monitoring is started with a non-admitted telemetry patient, the
printout type selection will be sent to the telemetry transmitter.
CAUTION
Users should be aware that all waveforms may be delayed
up to 2 seconds in the combination monitoring mode. If the
delay needs to be avoided, the combination monitoring mode
should be discontinued and all waveforms should be acquired
via the hard-wired bedside monitor.
CAUTION
Users should be aware of a possible time discrepancy
between the waveforms from the telemetry device and the
waveforms from a device hard-wired to the monitor. Users
should not consider these waveforms to be synchronous. If
absolute synchronicity is desired, the combination monitoring
mode should be discontinued and the ECG waveforms should
be acquired via the hard-wired bedside monitor.
Selecting the ECG source
NOTE
Not available with the NICU software package.
This setting is available if the telemetry license is enabled and the combination
monitoring mode
Monitor or Telemetry
has been enabled during configuration. The
monitor needs to be in the CARESCAPE Network.
1.
Select the HR parameter window.
2.
Select a source from the
ECG Source
list.
This list contains the monitor and available telemetry transmitter(s). When you
confirm the source selection with
Confirm
, the connection between the selected
transmitter and monitor will be established (telemetry transmitter selected), or the
patient is discharged from the telemetry transmitter (monitor selected).
Using the ECG measurement
The first three displayed ECG leads
You can choose the order of the ECG waveforms displayed in the ECG waveform area.
Lead selection depends on the type of ECG cable used.
The
ECG 1 Lead
,
ECG 2 Lead
, and
ECG 3 Lead
settings affect arrhythmia detection.
When
ECG 1 Lead
,
ECG 2 Lead
, or
ECG 3 Lead
are changed manually and the lead
becomes inactive due to a disconnection, the monitor looks to the ECG lead saved in
the patient profile. If
ECG 1 Lead
is not available, the monitor looks for lead II, then
lead I, and lastly lead III. Later, if the manually selected lead becomes available again,
the monitor will change back to this lead.
156
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...